FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 ARTICULATING

MDR report key: 6124002 · Received November 23, 2016

Report

Report Number
3005075853-2016-06673
Event Type
Malfunction
Date Received
November 23, 2016
Date of Event
November 16, 2016
Report Date
November 17, 2016
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K122797
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT/BATCH. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). BATCH # N9067W. THE DEVICE WAS RECEIVED WITH NO APPARENT DAMAGE. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). NO ISSUES WERE NOTED WITH OPENING AND CLOSING OF THE JAWS. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THE BATCH HISTORY RECORD WAS REVIEWED AND NO DEFECTS, NCR¿S OR PROTOCOLS RELATED TO THE COMPLAINT, WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC UNKNOWN PROCEDURE, THE JAWS BECAME NOT TO OPEN. THE BLADE TIP WAS NOT BROKEN OFF AND NO PIECES FELL INTO THE PATIENT. THE BLADE TIP WAS OPENED BY A HAND WITHOUT TISSUE DAMAGE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
775055 ENSEAL G2 ARTICULATING ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. NA N9080H

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR