ENSEAL G2 ARTICULATING
Report
- Report Number
- 3005075853-2016-06673
- Event Type
- Malfunction
- Date Received
- November 23, 2016
- Date of Event
- November 16, 2016
- Report Date
- November 17, 2016
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GEI
- PMA / PMN Number
- K122797
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WERE MET PRIOR TO THE RELEASE OF THIS LOT/BATCH. ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). BATCH # N9067W. THE DEVICE WAS RECEIVED WITH NO APPARENT DAMAGE. THE DEVICE WAS TESTED ON THE GENERATOR AND PASSED ALL FUNCTIONAL TESTING. THE ENERGY OUTPUT DELIVERED FROM THE DEVICE WAS VERIFIED. ALL THREE TONES WERE HEARD DURING FUNCTIONAL TESTING (TONE 1 IS HEARD WHEN THE ENERGY ACTIVATION BUTTON IS PRESSED; TONE 2 IS HEARD WHEN TISSUE IMPEDANCE THRESHOLD IS REACHED AND TONE 3 IS HEARD WHEN THE CYCLE IS COMPLETE). NO ISSUES WERE NOTED WITH OPENING AND CLOSING OF THE JAWS. THERE WERE NO ANOMALIES NOTED WITH THE FUNCTIONALITY OF THE DEVICE. THE BATCH HISTORY RECORD WAS REVIEWED AND NO DEFECTS, NCR¿S OR PROTOCOLS RELATED TO THE COMPLAINT, WERE FOUND DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC UNKNOWN PROCEDURE, THE JAWS BECAME NOT TO OPEN. THE BLADE TIP WAS NOT BROKEN OFF AND NO PIECES FELL INTO THE PATIENT. THE BLADE TIP WAS OPENED BY A HAND WITHOUT TISSUE DAMAGE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 775055 | ENSEAL G2 ARTICULATING | ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES | GEI | ETHICON ENDO-SURGERY, LLC. | NA | N9080H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR |