FDA Adverse Event Malfunction Summary report: N

CONQUEST 40 PTA BALLOON DILATATION CATHETER

MDR report key: 6123541 · Received November 22, 2016

Report

Report Number
2020394-2016-01104
Event Type
Malfunction
Date Received
November 22, 2016
Date of Event
October 14, 2016
Report Date
November 1, 2016
Manufacturer
BARD PERIPHERAL VASCULAR, INC.
Product Code
DQY
UDI-DI
00801741060557
PMA / PMN Number
K120660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING REVIEW: A MANUFACTURING REVIEW WAS CONDUCTED AND THERE WAS NOTHING FOUND TO INDICATE THERE WAS A MANUFACTURING RELATED CAUSE FOR THIS EVENT. THE LOT MET ALL RELEASE CRITERIA. VISUAL INSPECTION: THE BALLOON SIZE FOR THIS PRODUCT IS PRINTED ON THE BALLOON HUB OF THE CATHETER AND IDENTIFIED THE RETURNED SAMPLE AS A 8MM X 4CM BALLOON. NO ANOMALIES WERE NOTED TO THE DEVICE. FUNCTIONAL/PERFORMANCE EVALUATION: THE PATENCY OF THE GUIDEWIRE LUMEN WAS TESTED USING AN IN-HOUSE 0.035¿ GUIDEWIRE, AND IT PASSED WITHOUT ISSUE. THE BALLOON WAS CONNECTED TO AN IN-HOUSE INFLATION DEVICE, WITHOUT THE ONE WAY VALVE ATTACHED. AN ATTEMPT WAS THEN MADE TO INFLATE THE BALLOON. THE BALLOON WAS INFLATED TO RBP (40 ATM), AND THEN DEFLATED. THE BALLOON TOOK THE APPROPRIATE SHAPE UPON INFLATION AND WAS NOT ASYMMETRICAL IN APPEARANCE. THE DEFLATION TIME OF THE BALLOON WAS APPROXIMATELY 4.0 SECONDS, WHICH IS WITHIN THE SPECIFICATION OF 60 SECONDS. THE SAME TEST WAS REPEATED WITH THE INFLATION LUER ATTACHED, YIELDING THE SAME RESULTS. MEDICAL RECORDS REVIEW: MEDICAL RECORDS WERE NOT PROVIDED; THEREFORE, A REVIEW COULD NOT BE PERFORMED. IMAGE/PHOTO REVIEW: IMAGES/PHOTOS WERE NOT PROVIDED; THEREFORE, A REVIEW COULD NOT BE PERFORMED. CONCLUSION: THE INVESTIGATION WAS UNCONFIRMED FOR DEFLATION ISSUES, AS THE BALLOON WAS ABLE TO BE INFLATED AND DEFLATED WITHOUT ISSUE DURING LABORATORY TESTING. THE DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED BASED UPON AVAILABLE INFORMATION. IT WAS UNKNOWN WHETHER PATIENT AND/OR PROCEDURAL ISSUES CONTRIBUTED TO THE EVENT. IT SHOULD BE NOTED THAT PER THE REPORTED EVENT DETAILS, IT WAS STATED THAT THE DEVICE WAS INFLATED WITH A SYRINGE. THE IFU (INSTRUCTIONS FOR USE) STATES THAT THE USE OF A PRESSURE MONITORING DEVICE IS RECOMMENDED TO PREVENT OVER PRESSURIZATION. IT IS UNKNOWN WHETHER THE BALLOON/CATHETER WAS OVER PRESSURIZED OR THE USE OF A SYRINGE CONTRIBUTED TO THE EVENT. LABELING REVIEW: THE CURRENT IFU (INSTRUCTIONS FOR USE) STATES: WARNINGS: - DO NOT EXCEED THE RBP RECOMMENDED FOR THIS DEVICE. BALLOON RUPTURE MAY OCCUR IF THE RBP RATING IS EXCEEDED. TO PREVENT OVER PRESSURIZATION, USE OF A PRESSURE MONITORING DEVICE IS RECOMMENDED. - WHEN THE CATHETER IS EXPOSED TO THE VASCULAR SYSTEM, IT SHOULD BE MANIPULATED WHILE UNDER HIGH-QUALITY FLUOROSCOPIC OBSERVATION. DO NOT ADVANCE OR RETRACT THE CATHETER UNLESS THE BALLOON IS FULLY DEFLATED. IF RESISTANCE IS MET DURING MANIPULATION, DETERMINE THE CAUSE OF THE RESISTANCE BEFORE PROCEEDING. APPLYING EXCESSIVE FORCE TO THE CATHETER CAN RESULT IN TIP BREAKAGE OR BALLOON SEPARATION. PRECAUTIONS: - IF RESISTANCE IS FELT DURING POST PROCEDURE WITHDRAWAL OF THE CATHETER THROUGH THE INTRODUCER SHEATH, DETERMINE IF CONTRAST IS TRAPPED IN THE BALLOON WITH FLUOROSCOPY. IF CONTRAST IS PRESENT, PUSH THE BALLOON OUT OF THE SHEATH AND THEN COMPLETELY EVACUATE THE CONTRAST BEFORE PROCEEDING TO WITHDRAW THE BALLOON. - IF RESISTANCE IS STILL FELT DURING POST PROCEDURE WITHDRAWAL OF THE CATHETER, IT IS RECOMMENDED TO REMOVE THE BALLOON CATHETER AND GUIDEWIRE/INTRODUCER SHEATH AS A SINGLE UNIT. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Additional Manufacturer Narrative · 1

NO HOSPITAL/MEDICAL RECORDS OR MEDICAL IMAGES HAVE BEEN MADE AVAILABLE TO THE MANUFACTURER. AS THE LOT NUMBER FOR THE DEVICE WAS PROVIDED, A REVIEW OF THE DEVICE HISTORY RECORDS IS CURRENTLY BEING PERFORMED. THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INVESTIGATION OF THE REPORTED EVENT IS CURRENTLY UNDERWAY. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE ON A 50-59% STENOSIS OF A FOCAL LESION IN A VEIN, THE PTA BALLOON WAS ALLEGEDLY DIFFICULT TO DEFLATE. REPORTEDLY, THE ORIGINAL PTA BALLOON WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ANGIOPLASTY PROCEDURE ON A 50-59% STENOSIS OF A FOCAL LESION IN A VEIN, THE PTA BALLOON WAS ALLEGEDLY DIFFICULT TO DEFLATE. THE PTA BALLOON WAS EXCHANGED OVER THE GUIDEWIRE FOR ANOTHER THAT WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
773857 CONQUEST 40 PTA BALLOON DILATATION CATHETER PTA BALLOON DILATATION CATHETER DQY BARD PERIPHERAL VASCULAR, INC. REAU0299 00801741060557

Patients

Seq Age Sex Outcome Treatment
1 60 YR