PUMP IN STYLE ADVANCED - ON-THE-GO TOTE
Report
- Report Number
- 1419937-2016-00301
- Event Type
- Malfunction
- Date Received
- November 22, 2016
- Date of Event
- February 9, 2016
- Report Date
- November 22, 2016
- Manufacturer
- MEDELA, INC
- Product Code
- HGX
- PMA / PMN Number
- K031614
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
A REPLACEMENT PUMP WAS SENT TO THE CUSTOMER. THE PRODUCT WAS EVALUATED ON 10/26/2016, AND THE TRANSFORMER WAS DETERMINED TO BE BREACH BY QE. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING MONITORED FOR EFFECTIVENESS OF THE ABOVE MENTIONED CORRECTIVE ACTION.
THE CUSTOMER REPORTED TO CUSTOMER SERVICE ON (B)(6) 2016, THAT HER PUMP IN STYLE ADVANCED WILL NOT POWER ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 771033 | PUMP IN STYLE ADVANCED - ON-THE-GO TOTE | PUMP, BREAST, POWERED | HGX | MEDELA, INC | 57063/9207010 | REV M - 5112 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |