FDA Adverse Event Malfunction Summary report: N

X SERIES

MDR report key: 6121886 · Received November 22, 2016

Report

Report Number
1220908-2016-02911
Event Type
Malfunction
Date Received
November 22, 2016
Report Date
November 7, 2016
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K112432
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY RELAY ON THE ECG BOARD. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "DEFIB SHORT TEST FAILED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
771359 X SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION X SERIES NA

Patients

Seq Age Sex Outcome Treatment
1