FREESTYLE LITE
Report
- Report Number
- 2954323-2016-06881
- Event Type
- Malfunction
- Date Received
- November 22, 2016
- Date of Event
- November 10, 2016
- Report Date
- March 28, 2017
- Manufacturer
- ABBOTT DIABETES CARE INC
- Product Code
- NBW
- PMA / PMN Number
- K092602
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
ADDITIONAL INFORMATION: METER SERIAL (B)(4) HAS ALSO BEEN FILED UNDER MDR 2954323-2016-06917. THE REPORTED METER WAS RETURNED AND INVESTIGATED WITH RETAINED TEST STRIPS. THE METER WAS VISUALLY INSPECTED AND NO PHYSICAL DAMAGE WAS OBSERVED. METER POWERED ON WITH BUTTON DEPRESSION AND STRIP INSERTION. A BLANK SCREEN WAS NOT OBSERVED. THE COMPLAINT IS NOT CONFIRMED.
UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
THIS SERVES AS A CORRECTION REPORT. BRAND NAME WAS INCORRECTLY IDENTIFIED AS FREESTYLE FREEDOM LITE. THE CORRECT BRAND NAME FOR THE REPORTED METER SERIAL NUMBER IS FREESTYLE LITE. BRAND NAME HAS BEEN UPDATED TO REFLECT THE CORRECTION. IN ADDITION PMA/510K# HAS BEEN UPDATED TO K092602.
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 770999 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC | 71501-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |