FDA Adverse Event Malfunction Summary report: N

UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM

MDR report key: 6121017 · Received November 22, 2016

Report

Report Number
2523190-2016-00197
Event Type
Malfunction
Date Received
November 22, 2016
Report Date
November 8, 2016
Manufacturer
INTEGRA YORK, PA INC.
Product Code
FST
PMA / PMN Number
K071218
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON 11/30/16 INTEGRA INVESTIGATION COMPLETED. METHOD: FAILURE ANALYSIS, DEVICE HISTORY EVALUATION. RESULTS: FAILURE ANALYSIS - THERE WAS A UL PRO FUSED CABLE RETURNED BECAUSE THE END FITTING WAS BURNT BY THE MLX LIGHT SOURCE. THERE IS NO DEFICIENCY IDENTIFIED WITH THE CABLE. THE LIGHT SOURCE CAUSE THE DAMAGE TO THE CABLE. WITHOUT THE LIGHT SOURCE IT CANNOT BE DETERMINED WHY LIGHT SOURCE GENERATED ENOUGH HEAT TO DAMAGE THE CABLE. THE COMPLAINT CAN BE CONFIRMED, DEVICE HISTORY EVALUATION - NONCONFORMING PRODUCT REPORT / NONCONFORMING MATERIAL REPORT HISTORY: NONE. VARIANCE AUTHORIZATION / DEVIATION HISTORY: NONE. ENGINEERING CHANGE ORDER / MANUFACTURING CHANGE ORDER HISTORY: NONE. CORRECTIVE ACTION PREVENTIVE ACTION HISTORY: NONE. HEALTH HAZARD EVALUATION HISTORY: NONE. REPAIR HISTORY: N/A. CONCLUSION: THERE IS NO DEFICIENCY IDENTIFIED WITH THE CABLE. THE ROOT CAUSE CAN NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.

Description of Event or Problem · 1

CUSTOMER INITIALLY REPORTS CABLE END-FITTING BURNED BY MLX. ON (B)(6) 2016 CUSTOMER REPORTS ISSUE WAS FOUND DURING AN OPEN GALLBLADDER, ABOUT HALFWAY THROUGH THE PROCEDURE, NO HARM TO PATIENT. NO FURTHER INFORMATION AVAILABLE. #2 OF 2 RELATED DEVICES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
772143 UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM CABLES FST INTEGRA YORK, PA INC.

Patients

Seq Age Sex Outcome Treatment
1