UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM
Report
- Report Number
- 2523190-2016-00197
- Event Type
- Malfunction
- Date Received
- November 22, 2016
- Report Date
- November 8, 2016
- Manufacturer
- INTEGRA YORK, PA INC.
- Product Code
- FST
- PMA / PMN Number
- K071218
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
ON 11/30/16 INTEGRA INVESTIGATION COMPLETED. METHOD: FAILURE ANALYSIS, DEVICE HISTORY EVALUATION. RESULTS: FAILURE ANALYSIS - THERE WAS A UL PRO FUSED CABLE RETURNED BECAUSE THE END FITTING WAS BURNT BY THE MLX LIGHT SOURCE. THERE IS NO DEFICIENCY IDENTIFIED WITH THE CABLE. THE LIGHT SOURCE CAUSE THE DAMAGE TO THE CABLE. WITHOUT THE LIGHT SOURCE IT CANNOT BE DETERMINED WHY LIGHT SOURCE GENERATED ENOUGH HEAT TO DAMAGE THE CABLE. THE COMPLAINT CAN BE CONFIRMED, DEVICE HISTORY EVALUATION - NONCONFORMING PRODUCT REPORT / NONCONFORMING MATERIAL REPORT HISTORY: NONE. VARIANCE AUTHORIZATION / DEVIATION HISTORY: NONE. ENGINEERING CHANGE ORDER / MANUFACTURING CHANGE ORDER HISTORY: NONE. CORRECTIVE ACTION PREVENTIVE ACTION HISTORY: NONE. HEALTH HAZARD EVALUATION HISTORY: NONE. REPAIR HISTORY: N/A. CONCLUSION: THERE IS NO DEFICIENCY IDENTIFIED WITH THE CABLE. THE ROOT CAUSE CAN NOT BE DETERMINED.
TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
CUSTOMER INITIALLY REPORTS CABLE END-FITTING BURNED BY MLX. ON (B)(6) 2016 CUSTOMER REPORTS ISSUE WAS FOUND DURING AN OPEN GALLBLADDER, ABOUT HALFWAY THROUGH THE PROCEDURE, NO HARM TO PATIENT. NO FURTHER INFORMATION AVAILABLE. #2 OF 2 RELATED DEVICES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 772143 | UL PRO FUSED HEADLIGHT CABLE 9FT, 275CM | CABLES | FST | INTEGRA YORK, PA INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |