FDA Adverse Event Malfunction Summary report: N

NCP BIPOLAR LEAD

MDR report key: 612090 · Received November 5, 2004

Report

Report Number
1644487-2004-00963
Event Type
Malfunction
Date Received
November 5, 2004
Report Date
October 7, 2004
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

PT UNDERWENT NCP SYSTEM REPLACEMENT DUE TO SUSPECTED LEAD BREAK. IT WAS ALSO REPORTED THAT DEVICE DIAGNOSTIC TESTING PRIOR TO NCP SYSTEM REPLACEMENT RESULTED IN HIGH LEAD IMPEDANCE READING (DC-DC CODE 7) INDICATING POSSIBLE DEVICE MALFUNCTION. DURING THE SURGERY, SURGEON IDENTIFIED THAT THE LEAD WAS BROKEN AND WAS COVERED WITH A HARD DRY YELLOW-CLEAR SUBSTANCE. FURTHER FOLLOW-UP REVEALED THAT THE HIGH LEAD IMPEDANCE READING WAS RESOLVED WITH NCP SYSTEM REPLACEMENT. NEUROLOGIST INDICATED THAT THE PT IS DOING GOOD SINCE GENERATOR REPLACEMENT SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NCP BIPOLAR LEAD BIPOLAR LEAD LYJ CYBERONICS, INC. 300-20 27779C

Patients

Seq Age Sex Outcome Treatment
1 46 YR