ITREVIA 7 HF-T QP DF4 IS4
Report
- Report Number
- 1028232-2016-04662
- Event Type
- Death
- Date Received
- November 22, 2016
- Date of Event
- September 30, 2016
- Report Date
- November 14, 2016
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NIK
- PMA / PMN Number
- P050023
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- OTHER
Narratives
WE RECEIVED YOUR EVENT DESCRIPTION FOR THE ABOVE MENTIONED DEVICE AND WOULD LIKE TO THANK YOU FOR SUPPORTING OUR POST-MARKET SURVEILLANCE. AS OF TODAY, THE MEDICAL DEVICE IS NOT AVAILABLE FOR ANALYSIS, THEREFORE THE DEVICE ITSELF COULD NOT BE INVESTIGATED. THE INFORMATION YOU PROVIDED HAS BEEN ENTERED INTO OUR QUALITY SYSTEM AS A COMPLAINT. THESE TYPES OF COMPLAINTS ARE USED TO EVALUATE SYSTEMS AND DEVICE PERFORMANCE THROUGHOUT OUR ORGANIZATION AND HELP TO MAINTAIN AND IMPROVE THE PERFORMANCE OF OUR DEVICES. SHOULD ADDITIONAL RELEVANT INFORMATION OR THE DEVICE ITSELF BECOME AVAILABLE, THE INVESTIGATION WILL BE UPDATED.
(B)(4).
THE DEVICE WAS RECEIVED AND ANALYZED. THE MEMORY CONTENT OF THE ICD WAS INSPECTED, SHOWING A NORMAL DEVICE FUNCTION WHILE IMPLANTED AND IN SERVICE. IN A NEXT STEP, A SENSING TEST WAS PERFORMED AND THE DEVICE SENSED THE ATTACHED HEART SIGNALS FREE OF NOISE. THE ICD WAS SUBJECTED TO AN ELECTRICAL ANALYSIS. THE ABILITY OF THE DEVICE TO DELIVER THERAPIES WAS VERIFIED. THE ANTI-BRADYCARDIA PACING PULSES PROVED TO BE NORMAL AND IN AMPLITUDE AND FREQUENCY AS PROGRAMMED. A FIBRILLATION SIGNAL WAS APPLIED AND THE DEVICE DELIVERED A DEFIBRILLATION SHOCK AS SPECIFIED, DOCUMENTING A CORRECT SENSING AND SHOCK DELIVERY. IN PARTICULAR, THE SPECIFIED ENERGY LEVEL WAS REACHED. IN CONCLUSION, THE DEVICE PROVED TO BE FULLY FUNCTIONAL. THE ANALYSIS OF THE MEMORY CONTENT SHOWED A NORMAL DEVICE BEHAVIOR WHILE THE DEVICE WAS IMPLANTED AND IN SERVICE. THERE WAS NO INDICATION OF A DEVICE MALFUNCTION.
WE WERE NOTIFIED BY THE PATIENTS FAMILY THAT THIS PATIENT EXPIRED ONE WEEK AFTER THE ICD SYSTEM WAS IMPLANTED. THE FAMILY NOTED THAT THIS DEVICE DELIVERED 3 SHOCKS THE DAY AFTER IMPLANT AND THAT THERAPY WAS TURNED OFF 3 DAYS AFTER SURGERY BECAUSE THE MEDICATION THE PATIENT WAS ON AND THE DEVICE WERE TOO MUCH FOR THE PATIENTS BODY. THE FAMILY BELIEVES THAT THIS DEVICE WAS EXPLANTED BEFORE THE PATIENT WAS CREMATED. THIS DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 773359 | ITREVIA 7 HF-T QP DF4 IS4 | CRT-D | NIK | BIOTRONIK SE & CO. KG | 401662 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Death |