FDA Adverse Event Malfunction Summary report: N

ENDO LINEAR CUTTER

MDR report key: 6120607 · Received November 22, 2016

Report

Report Number
3005075853-2016-06661
Event Type
Malfunction
Date Received
November 22, 2016
Date of Event
October 21, 2016
Report Date
October 26, 2016
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BATCH # N5637D. THE ANALYSIS SHOWED THAT THE ATS45 DEVICE WAS RECEIVED WITH NO APPARENT DAMAGE AND WITH A TR45W CARTRIDGE LOADED IN THE DEVICE. THE RELOAD WAS RECEIVED PARTIALLY FIRED 1/10 AND WITH THE RELOAD LOCK OUT SPRING DAMAGED. THE DAMAGE TO THE RELOAD LOCKOUT SPRING IS CONSISTENT WITH DAMAGE OBSERVED WHEN THE FIRING CYCLE IS STARTED, INTERRUPTED, RELEASED, AND RESTARTED. WHEN FIRING THE DEVICE MAKE SURE THAT THE FIRING STROKE IS COMPLETED. DO NOT PARTIALLY FIRE THE DEVICE. FIRE THE DEVICE BY SQUEEZING THE FIRING TRIGGER COMPLETELY UNTIL IT RESTS ON THE CLOSING TRIGGER. ONCE THE FIRING CYCLE HAS BEEN INITIATED, IT MUST BE COMPLETED. IF RE-INITIATION OF FIRING IS RESUMED, THE DEVICE WILL LOCKOUT. FIRING THROUGH THE LOCKOUT MECHANISM WILL BREAK THE DEVICE. PLEASE REFERENCE THE INSTRUCTION FOR USE FOR MORE INFORMATION. THE RETURNED DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. THE BATCH HISTORY RECORD WAS REVIEWED AND NO DEFECTS, NCR¿S OR PROTOCOLS RELATED TO THE COMPLAINT, WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THAT THE STAPLER LOADED AND TESTED OKAY. THE DEVICE CLOSED ON TISSUE, BUT DID NOT FIRE. IT IS UNKNOWN HOW THE DEVICE WAS REMOVED. THE PROCEDURE WAS COMPLETED USING ANOTHER LIKE DEVICE OF THE SAME PRODUCT CODE. NO OTHER INFORMATION WAS AVAILABLE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
771339 ENDO LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA N4LZ0L

Patients

Seq Age Sex Outcome Treatment
1