LEAD MODEL 302
Report
- Report Number
- 1644487-2016-02702
- Event Type
- Malfunction
- Date Received
- November 21, 2016
- Date of Event
- August 1, 2016
- Report Date
- June 9, 2017
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
IT WAS REPORTED THAT THE PATIENT'S UNDERWENT FULL REVISION OF GENERATOR AND LEAD DUE TO HIGH IMPEDANCE. THE SUSPECT PRODUCT WAS RECEIVED FOR ANALYSIS. NO FURTHER RELEVANT INFORMATION HAS BEEN RECEIVED TO DATE.
ANALYSIS WAS COMPLETED ON THE GENERATOR AND LEAD. A COMPREHENSIVE AUTOMATED ELECTRICAL EVALUATION SHOWED THAT THE PULSE GENERATOR PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS VARIOUS ELECTRICAL LOADS WERE ATTACHED TO THE PULSE GENERATOR AND RESULTS OF DIAGNOSTIC TESTS DEMONSTRATE THAT ACCURATE RESISTANCE MEASUREMENTS WERE OBTAINED IN ALL INSTANCES. NO ANOMALIES WERE FOUND, THE INTERNAL DATA OF THE GENERATOR WAS REVIEWED AND FOUND THAT THE LAST 25% CHANGE IN IMPEDANCE OCCURRED ON (B)(6) 2017 FROM IMPEDANCE WITHIN NORMAL LIMITS (4281 OHMS) TO HIGH IMPEDANCE (5633 OHMS). THIS SUGGESTS AN INTERMITTENT LEAD FRACTURE WITH FLUCTUATING HIGH IMPEDANCE AS HIGH IMPEDANCE WAS DETECTED PRIOR TO (B)(6) 2017. PRODUCT ANALYSIS WAS COMPLETED ON THE LEAD. THE LEAD ASSEMBLY WAS RETURNED IN TWO PORTIONS. THE ELECTRODE ARRAY WAS NOT RETURNED, SO A COMPLETE ANALYSIS COULD NOT BE PERFORMED ON THE PRODUCT. SET-SCREW MARKS WERE OBSERVED ON THE LEAD PIN, WHICH INDICATES THAT AT ONE POINT IN TIME THERE WAS A GOOD ELECTRICAL AND MECHANICAL CONNECTION BETWEEN THE LEAD AND GENERATOR. THERE WERE MULTIPLE INSTANCES OF ABRADED OPENINGS OF THE INNER TUBING AND OF THE INNER AND OUTER TUBING. PORTIONS OF THE QUADFILAR COIL WERE EXPOSED. THERE WERE BODILY FLUIDS FOUND IN BOTH THE INNER AND OUTER TUBING THAT WERE ATTRIBUTED TO THESE OPENINGS. THE CAUSE OF THESE ABRADED OPENING WAS DETERMINED TO BE WEAR. A LEAD FRACTURE WAS NIT FOUND ON THE RETURNED PORTIONS. BEYOND THE ABRADED OPENINGS, THE CONDITION OF THE LEAD WAS TYPICAL OF THAT OF LEADS AFTER EXPLANT. NO FURTHER ANOMALIES FOUND. NO FURTHER RELEVANT INFORMATION HAS BEEN RECEIVED TO DATE.
THE SURGEON REPORTED FINDING HIGH IMPEDANCE ON THE PATIENT'S GENERATOR DURING A VISIT ON (B)(6) 2016. THERE WAS NO KNOWN TRAUMA, BUT IT WAS STATED THAT TRAUMA WAS "CERTAINLY POSSIBLE GIVEN HIS CONDITION". NO REVISION SURGERY FOR THE HIGH IMPEDANCE HAS OCCURRED TO DATE. NO ADDITIONAL RELEVANT INFORMATION HAS BEEN RECEIVED TO DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 770294 | LEAD MODEL 302 | LEAD | LYJ | CYBERONICS, INC. | 302-20 | 5986 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR |