FDA Adverse Event
Malfunction
Summary report: N
TRIAGE PROFILER SOB PANEL
MDR report key: 6119121
·
Received November 21, 2016
Report
- Report Number
- 2027969-2016-00686
- Event Type
- Malfunction
- Date Received
- November 21, 2016
- Date of Event
- October 26, 2016
- Report Date
- October 27, 2016
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- NBC
- PMA / PMN Number
- K080269
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION CONCLUSION: CUSTOMER'S COMPLAINT WAS NOT REPLICATED WITH IN-HOUSE TESTING OF RETAIN LOT W62069. NO ISSUES WITH CKMB RECOVERY OBSERVED. REVIEWED THE BATCH RECORD FOR LOT W62069. LOT MET ALL OUS FINAL RELEASE SPECIFICATIONS. UNABLE TO DETERMINE ROOT CAUSE OF COMPLAINT. PRODUCT DEFICIENCY WAS NOT ESTABLISHED.
Description of Event or Problem · 1
EVENT OCCURRED IN (B)(6). CUSTOMER ALLEGING TRIAGE PROFILER SOB PANEL HAD DISCREPANT RESULTS WITH CK-MB IN COMPARISON TO THE RESULTS FROM THE HOSPITAL LABORATORY. TRIAGE CK MB RESULT 33,3 ( SHOULD BE WITHIN 0-4,3) AND IN THE HOSPITAL AFTER VENOUS BLOOD SAMPLE IT WAS 17,6 AND IT SHOULD BE LESS THAN 24. NO REPORTED ADVERSE PATIENT SEQUELA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 767483 | TRIAGE PROFILER SOB PANEL | CARDIAC MARKER TEST | NBC | ALERE SAN DIEGO, INC. | 97300EU | W62069B |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | TRIAGE METERPRO INTERNATIONAL SERIAL #(B)(4) |