DENALI JUGULAR SYSTEM

MDR report key: 6118634 · Received November 21, 2016

Report

Report Number
2020394-2016-01093
Event Type
Malfunction
Date Received
November 21, 2016
Date of Event
October 21, 2016
Report Date
October 25, 2016
Manufacturer
BARD PERIPHERAL VASCULAR, INC.
Product Code
DTK
UDI-DI
00801741040795
PMA / PMN Number
K130366
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MANUFACTURING REVIEW: THE DEVICE HISTORY RECORDS HAVE BEEN REVIEWED WITH SPECIAL ATTENTION TO THE RAW MATERIALS, SUBASSEMBLIES, MANUFACTURING PROCESS AND QUALITY CONTROL TESTING. THIS LOT MET ALL RELEASE CRITERIA. THERE WAS NOTHING FOUND TO INDICATE THERE WAS A MANUFACTURING RELATED CAUSE FOR THIS EVENT. VISUAL INSPECTION: ALL 6 ARMS AND 6 LEGS WERE PRESENT AND INTACT. NO LIMBS WERE CROSSED. NO ANOMALIES TO THE FILTER WERE NOTED. DIMENSIONAL EVALUATION: ALL DIMENSIONAL MEASUREMENTS MET THE REQUIRED SPECIFICATIONS. MEDICAL RECORDS REVIEW: MEDICAL RECORDS WERE NOT PROVIDED FOR REVIEW. IMAGE/PHOTO REVIEW: IMAGES/PHOTOS WERE NOT PROVIDED FOR REVIEW. CONCLUSION: THE DEVICE WAS RETURNED. NO CROSSED OR BENT LIMBS WERE IDENTIFIED ON THE PHYSICAL SAMPLE. BASED ON THE DIMENSIONAL EVALUATION THE INVESTIGATION WAS INCONCLUSIVE FOR FAILURE TO EXPAND. BASED UPON THE AVAILABLE INFORMATION, THE DEFINITIVE ROOT CAUSE FOR THIS EVENT WAS UNKNOWN AS IMAGES WERE NOT PROVIDED. LABELING REVIEW: THE CURRENT IFU (INSTRUCTIONS FOR USE) STATES: WARNINGS: - DELIVERY OF THE DENALI FILTER THROUGH THE INTRODUCER SHEATH IS ADVANCE ONLY. RETRACTION AND TWISTING OF THE PUSHER DURING DELIVERY COULD RESULT IN DISLODGEMENT OF THE FILTER, CROSSING OF FILTER LEGS OR ARMS, AND COULD PREVENT THE FILTER FROM FURTHER ADVANCEMENT WITHIN THE INTRODUCER SHEATH. PRECAUTIONS: - CARE SHOULD BE TAKEN TO ENSURE THE CONNECTION BETWEEN THE INTRODUCER SHEATH HUB AND THE FILTER STORAGE TUBE IS TIGHT; HOWEVER, THE USE OF EXCESSIVE FORCE WHICH CAN CAUSE SLIPPAGE OF THE THREADS AND/OR BREAKAGE OF THE HUB SHOULD BE AVOIDED. - DO NOT DELIVER THE FILTER BY PUSHING IT BEYOND THE END OF THE INTRODUCER SHEATH. TO ACHIEVE PROPER PLACEMENT, UNSHEATH THE STATIONARY FILTER BY WITHDRAWING THE INTRODUCER SHEATH. DO NOT TWIST THE PUSHER AT ANYTIME DURING THIS PROCEDURE. POTENTIAL COMPLICATIONS: - FAILURE OF FILTER EXPANSION/INCOMPLETE EXPANSION. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Additional Manufacturer Narrative · 1

NO MEDICAL RECORDS OR NO IMAGES HAVE BEEN MADE AVAILABLE TO THE MANUFACTURER. AS THE LOT NUMBER FOR THE DEVICE WAS PROVIDED, A REVIEW OF THE DEVICE HISTORY RECORDS IS CURRENTLY BEING PERFORMED. THE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. THE INVESTIGATION OF THE REPORTED EVENT IS CURRENTLY UNDERWAY. THE INFORMATION PROVIDED BY BARD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BARD.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING DEPLOYMENT OF A VENA CAVA FILTER, THE FILTER LEGS ALLEGEDLY CROSSED AND WOULD NOT OPEN. IT WAS FURTHER REPORTED THAT THE FILTER WAS CAPTURED AND RETRIEVED AND ANOTHER VENA CAVA FILTER WAS SUCCESSFULLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING DEPLOYMENT OF A VENA CAVA FILTER, THE FILTER LEGS ALLEGEDLY CROSSED AND WOULD NOT OPEN. IT WAS FURTHER REPORTED THAT THE FILTER WAS CAPTURED AND RETRIEVED AND ANOTHER VENA CAVA FILTER WAS SUCCESSFULLY DEPLOYED. THERE WAS NO REPORTED PATIENT INJURY.