PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2016-36879
- Event Type
- Malfunction
- Date Received
- November 21, 2016
- Date of Event
- March 11, 2014
- Report Date
- March 11, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PATIENT
Narratives
INSULIN PUMP RECEIVED WITH BUTTON ERROR ALARM DUE TO FLATTENED BUTTON DOME SWITCH. UNABLE TO PERFORM DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST DUE TO BUTTON ERROR ALARM. DEVICE HAD SCRATCHED DISPLAY WINDOW AND CRACKED DISPLAY WINDOW CORNER. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. (B)(4).
CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROFESSIONALS' INSTRUCTIONS. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL BE REPLACED. CUSTOMER AGREED TO RETURN THE DEVICE FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 769790 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754CAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |