INFUSOMAT®
Report
- Report Number
- 9614279-2016-00096
- Event Type
- Malfunction
- Date Received
- November 18, 2016
- Date of Event
- November 26, 2014
- Report Date
- December 1, 2014
- Manufacturer
- B. BRAUN DOMINICAN REPUBLIC INC.
- Product Code
- FPA
- UDI-DI
- 04046964294570
- PMA / PMN Number
- K142036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT # (B)(4). THE ACTUAL DEVICE INVOLVED IN THE REPORTED INCIDENT WAS NOT RETURNED FOR EVALUATION. WITHOUT THE ACTUAL SAMPLE A THOROUGH SAMPLE ANALYSIS COULD NOT BE PERFORMED AND NO SPECIFIC CONCLUSIONS CAN BE DRAWN. HOWEVER, PICTURES PROVIDED BY THE FACILITY DID CONFIRM THAT BLOOD WAS LEAKING NEAR THE BONDED JOINT OF THE DISTAL END OF THE PUMP SEGMENT. REVIEW OF THE DISCREPANCY MANAGEMENT SYSTEM DATABASE PERFORMED FOR THE REPORTED LOT NUMBER DID NOT REVEAL ANY ABNORMALITIES OR NONCONFORMANCES OF THIS NATURE. NO ADVERSE QUALITY TRENDS OF THIS NATURE WERE IDENTIFIED DURING THE COMPLAINT REVIEW PROCESS FOR THE REPORTED CATALOG NUMBER OR LOT NUMBER. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE A FOLLOW-UP REPORT WILL BE FILED. NOTE: THIS CASE IS BEING FILED RETROSPECTIVELY AS A RESULT OF A REVIEW OF RECENT CUSTOMER COMPLAINT INFORMATION. BASED ON ADDITIONAL INFORMATION AND DETAILS PROVIDED IN ANOTHER COMPLAINT CASE, IT WAS DETERMINED THAT THIS CASE IS REPORTABLE IN ACCORDANCE WITH THE REQUIREMENTS OF 21 CFR 803. B. BRAUN HAS CONDUCTED A RETROSPECTIVE REVIEW FOR ALL COMPLAINTS OF A SIMILAR NATURE IN ACCORDANCE WITH INTERNAL PROCEDURE (B)(4).
AS REPORTED BY USER FACILITY: BLOOD TUBING LEAKED AT THE PUMP SEGMENT SECTION NEAR THE WHITE PART.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 763113 | INFUSOMAT® | SET, ADMINISTRATION, INTRA | FPA | B. BRAUN DOMINICAN REPUBLIC INC. | 0061381033 | 04046964294570 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |