FDA Adverse Event Malfunction Summary report: N

AIM MEDICAL LIGHTS

MDR report key: 6114183 · Received November 18, 2016

Report

Report Number
3009542956-2016-00005
Event Type
Malfunction
Date Received
November 18, 2016
Date of Event
October 27, 2016
Report Date
November 17, 2016
Manufacturer
PHILIPS BURTON
Product Code
KZF
PMA / PMN Number
N/A
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CUSTOMER REPORTED THAT AIM LIGHT HEAD BECAME DETACHED. THE STOP HAD BEEN OVER-TORQUED AND SAFETY LOCKING KEY CRACKED THE COLLAR. NO INJURIES REPORTED. IT WAS FOUND THAT THERE ARE SEVEN (7) COMPLAINTS RELATED TO THE CRACKED COLLARS THAT RESULTED IN THE LIGHT HEAD DETACHED FROM THE SPRING ARM. NO INJURIES REPORTED AMONG THE SEVEN COMPLAINTS. UPON EXAMINATION, IT WAS DISCOVERED THAT THE PLASTIC COLLARS USED TO SECURE THE SPRING ARM WERE CRACKING DUE TO OVER-TORQUED DURING INSTALLATION WHICH RESULTED IN THE LIGHT HEAD TO SEPARATE FROM THE SPRING ARM. IN ORDER TO PREVENT THIS ISSUE FROM HAPPENING, PHILIPS BURTON IS REPLACING ALL PLASTIC COLLARS WITH NEW ALUMINUM COLLARS IN THE FIELD. FIELD ACTION DECISION WAS TAKEN TO PROCEED WITH THE FIELD CORRECTION.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT AIM LIGHT HEAD BECAME DETACHED. THE STOP HAD BEEN OVER-TORQUED AND SAFETY LOCKING KEY CRACKED THE COLLAR. NO INJURIES REPORTED. IT WAS FOUND THAT THERE ARE SEVEN (7) COMPLAINTS RELATED TO THE CRACKED COLLARS THAT RESULTED IN THE LIGHT HEAD DETACHED FROM THE SPRING ARM. NO INJURIES REPORTED AMONG THE SEVEN COMPLAINTS. UPON EXAMINATION, IT WAS DISCOVERED THAT THE PLASTIC COLLARS USED TO SECURE THE SPRING ARM WERE CRACKING DUE TO OVER-TORQUED DURING INSTALLATION WHICH RESULTED IN THE LIGHT HEAD TO SEPARATE FROM THE SPRING ARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764962 AIM MEDICAL LIGHTS KZF PHILIPS BURTON ALEDSC N/A