FDA Adverse Event Injury Summary report: N

LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI)

MDR report key: 6113793 · Received November 18, 2016

Report

Report Number
2023950-2016-00167
Event Type
Injury
Date Received
November 18, 2016
Date of Event
May 6, 2016
Report Date
August 25, 2016
Manufacturer
ZEST ANCHORS, LLC
Product Code
DZE
PMA / PMN Number
K120198
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

USER DOCUMENTATION (TECHNIQUE MANUAL, PN L8019-TM) SPECIFIES THAT IMMEDIATE LOADING IS SUITABLE ONLY IF SUFFICIENT PRIMARY STABILITY OF THE IMPLANT IS ACHIEVED AT THE TIME OF PLACEMENT. ALSO, THE RECOMMENDED IMPLANT INSERTION TORQUE AND IMPLANT ATTACHMENT TORQUE IS 30N-CM. IF THE IMPLANT PLACEMENT TORQUE IS LESS THAN 30 N-CM , THEN THE ATTACHMENTS SHOULD ONLY BE HAND TIGHTENED. THE IMPLANT INSERTION TORQUE SHOULD NOT EXCEED 70 N-CM; IF 70 N-CM IS REACHED PRIOR TO FULL SEATING, THE IMPLANT SHOULD BE REMOVED AND THE OSTEOTOMY SHOULD BE ENLARGED. CLINICIAN STATED THAT A LODI IMPLANT FAILED TO OSSEOINTEGRATE. PATIENT HAS LOW DENSITY BONE. IMPLANT WAS NOT IMMEDIATELY LOADED. THE CLINICIAN DID NOT PROVIDE THE FOLLOWING INFORMATION: AMOUNT OF TORQUE USED ON ABUTMENT & IMPLANT. WHETHER PRIMARY STABILITY WAS ACHIEVED. FAILURE TO OSSEOINTEGRATE IS A WELL-DOCUMENTED INHERENT RISK OF DENTAL IMPLANTS. IN THE MAJORITY OF CASES WHERE AN IMPLANT FAILS TO INTEGRATE WITH THE BONE AND IS REJECTED BY THE BODY, THE CAUSE IS UNKNOWN. THE RECORDS MANAGEMENT DATABASE OF EACH OF THE ORDERED IMPLANTS INDICATES THAT THE IMPLANTS MET THE SPECIFICATIONS AND WERE APPROVED FOR PRODUCT RELEASE. ANY ISSUES WERE SUCCESSFULLY RESOLVED PRIOR TO THE RELEASE OF THE IMPLANTS. NO FURTHER ACTION IS REQUIRED. REFER TO PER (B)(4).

Description of Event or Problem · 1

LODI IMPLANT FAILED TO OSSEOINTEGRATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762860 LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI) ENDOSSEOUS DENTAL IMPLANT DZE ZEST ANCHORS, LLC 07455 I10K0

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention