FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 611326 · Received July 14, 2004

Report

Report Number
2243072-2004-00374
Event Type
Other
Date Received
July 14, 2004
Date of Event
June 16, 2004
Report Date
July 14, 2004
Manufacturer
BECTON DICKINSON
Product Code
NBW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT VERY LOW BLOOD SUGAR RESULTS WITH THEIR PLINK METER. METER NOT REFLECTING HOW THEY FEEL; TESTING TO ANOTHER METER. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (167) VS. BD READNGS (26, 118, 162) YIELDED DATA POINTS IN THE C RANGE OF THE GRID, OUTSIDE ACCEPTABLE LIMITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 3064100

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other