O-ARM 1000 IMAGING SYSTEM
Report
- Report Number
- 1723170-2016-03582
- Event Type
- Malfunction
- Date Received
- November 18, 2016
- Date of Event
- January 14, 2016
- Report Date
- November 18, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
PATIENT INFORMATION WAS UNAVAILABLE FROM THE SITE. A SITE REPRESENTATIVE REPORTED THAT, WHILE IN A SPINAL FUSION PROCEDURE, BOTH THE IMAGE ACQUISITION SYSTEM (IAS) AND MOBILE VIEW STATION (MVS) STAYED IN STAND-ALONE MODE. AFTER A SYSTEM REBOOT OF BOTH COMPONENTS, CONNECTIVITY WAS REESTABLISHED AND THE CASE WAS CONTINUED. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME. ONSITE INVESTIGATION DISCOVERED MINOR DAMAGE DONE THE MVS UMBILICAL CABLE. REPLACEMENT OF THE CABLE RESOLVED THE ISSUE. NO FURTHER ISSUES WERE REPORTED. ANALYSIS OF THE RETURNED CABLE COULD NOT CONFIRM ANY FAILURE AS IT PASSED BENCH TESTING AND FUNCTIONED WITHOUT PROBLEMS. A FULL IMAGING SYSTEM CHECK-OUT WAS COMPLETED AND ALL TESTS PASSED. FULL SYSTEM FUNCTIONALITY WAS CONFIRMED AND THE SYSTEM WAS RETURNED TO SERVICE. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA NEW ENGLAND DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 FDA-483 FEI: 3004785967. THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
A SITE REPRESENTATIVE REPORTED THAT, WHILE IN A SPINAL FUSION PROCEDURE, BOTH THE IMAGE ACQUISITION SYSTEM (IAS) AND MOBILE VIEW STATION (MVS) STAYED IN STAND-ALONE MODE. AFTER A SYSTEM REBOOT OF BOTH COMPONENTS, CONNECTIVITY WAS REESTABLISHED AND THE CASE WAS CONTINUED. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 765687 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |