FDA Adverse Event Malfunction Summary report: N

ACCEL

MDR report key: 611262 · Received May 27, 2005

Report

Report Number
1034973-2005-00003
Event Type
Malfunction
Date Received
May 27, 2005
Date of Event
April 27, 2005
Report Date
April 28, 2005
Manufacturer
CTI PET SYSTEMS, INC.
Product Code
KPS
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING TRANSIT OF A MOBILE PET SYSTEM, THE PATIENT HANDLING SYSTEM (PHS) COLLAPSED. CPS HAS INVESTIGATED THIS PROBLEM AND FOUND THAT THE SUPPORT SCREWS ON THE PHS VERTICAL MOTOR MOUNT CAN BECOME LOOSE OVER TIME DUE TO INCREASED VIBRATION. THIS PROCESS THEN CAUSES STRESS AND BREAKAGE IN OTHER MOUNTING SCREWS. TWO BED SUPPORTS ARE REQUIRED TO BE USED WHILE THE SYSTEM IS IN TRANSIT. THIS REDUCES STRESS AND VIBRATION ON THE PHS WHILE THE MOBILE COACH IS IN TRANSIT. THIS MOBILE PROVIDER WAS NOT USING THE REQUIRED BED SUPPORTS, WHICH LIKELY CONTRIBUTED TO THIS PROBLEM. NO INJUIRIES HAVE BEEN REPORTED TO CPS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCEL PET SCANNER KPS CTI PET SYSTEMS, INC. 3600211-00 NA

Patients

Seq Age Sex Outcome Treatment
1 NA Other