FDA Adverse Event
Malfunction
Summary report: N
ACCEL
MDR report key: 611262
·
Received May 27, 2005
Report
- Report Number
- 1034973-2005-00003
- Event Type
- Malfunction
- Date Received
- May 27, 2005
- Date of Event
- April 27, 2005
- Report Date
- April 28, 2005
- Manufacturer
- CTI PET SYSTEMS, INC.
- Product Code
- KPS
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
DURING TRANSIT OF A MOBILE PET SYSTEM, THE PATIENT HANDLING SYSTEM (PHS) COLLAPSED. CPS HAS INVESTIGATED THIS PROBLEM AND FOUND THAT THE SUPPORT SCREWS ON THE PHS VERTICAL MOTOR MOUNT CAN BECOME LOOSE OVER TIME DUE TO INCREASED VIBRATION. THIS PROCESS THEN CAUSES STRESS AND BREAKAGE IN OTHER MOUNTING SCREWS. TWO BED SUPPORTS ARE REQUIRED TO BE USED WHILE THE SYSTEM IS IN TRANSIT. THIS REDUCES STRESS AND VIBRATION ON THE PHS WHILE THE MOBILE COACH IS IN TRANSIT. THIS MOBILE PROVIDER WAS NOT USING THE REQUIRED BED SUPPORTS, WHICH LIKELY CONTRIBUTED TO THIS PROBLEM. NO INJUIRIES HAVE BEEN REPORTED TO CPS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCEL | PET SCANNER | KPS | CTI PET SYSTEMS, INC. | 3600211-00 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |