VITROS CHEMISTRY PRODUCTS LAC SLIDES
Report
- Report Number
- 1319809-2016-00102
- Event Type
- Malfunction
- Date Received
- November 17, 2016
- Date of Event
- October 25, 2016
- Report Date
- November 17, 2016
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- KHP
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE INVESTIGATION DETERMINED THAT HIGHER AND LOWER THAN EXPECTED VITROS LAC QUALITY CONTROL RESULTS WERE OBTAINED FROM VITROS CONTROL FLUIDS USING TWO DIFFERENT LOTS OF VITROS LAC SLIDES ON A VITROS 350 CHEMISTRY SYSTEM. A DEFINITIVE ASSIGNABLE CAUSE COULD NOT BE DETERMINED. THERE IS NO EVIDENCE THAT A VITROS 350 SYSTEM OCCURRED. THE UNACCEPTABLE HISTORICAL VITROS LAC QUALITY CONTROL RESULTS OBTAINED FROM THE CUSTOMER INDICATE THAT THE VITROS LAC LOT 3530-0092-3271 WAS NOT PERFORMING AS INTENDED. HOWEVER, IT IS SUSPECTED THAT THE CUSTOMER IS NOT FOLLOWING CORRECT PROTOCOL WHEN RECONSTITUTING THE CALIBRATOR AND CONTROL FLUIDS BUT COULD NOT BE CONFIRMED.
THE CUSTOMER OBSERVED HIGHER AND LOWER THAN EXPECTED VITROS LAC QUALITY CONTROL RESULTS FROM VITROS CONTROL FLUIDS USING TWO DIFFERENT LOTS OF VITROS LAC SLIDES ON A VITROS 350 CHEMISTRY SYSTEM. VITROS LAC LOT 3530-0092-3271: PV2 LOT K4852 VITROS LAC RESULT 2.58*, 2.64*, 2.61*, 2.62*, 2.37*, 2.58*, 2.60*, 2.59*, 2.60*, 2.56*, 2.64*, 2.62*, 2.58*, 2.60*, AND 2.61* MMOL/L ,VERSUS THE EXPECTED VITROS LAC RESULT 3.96 MMOL/L; PV2 LOT M5007 VITROS LAC RESULTS 5.09*, 2.54* AND 5.19* MMOL/L VERSUS THE EXPECTED VITROS LAC RESULT 3.76 MMOL/L. VITROS LAC LOT 3531-0093-5063: PV1 LOT L5005 VITROS LAC RESULT 2.13* MMOL/L VERSUS THE EXPECTED VITROS LAC RESULT 1.39 MMOL/L; PV2 LOT M5007 VITROS LAC RESULT 5.67* MMOL/L VERSUS THE EXPECTED VITROS LAC RESULT 3.33 MMOL/L. BIASED RESULTS OF THE DIRECTION AND MAGNITUDE OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION. ALTHOUGH THERE WERE NO REPORTS OF AFFECTED PATIENT SAMPLE RESULTS, IT CANNOT BE CONFIRMED THAT PATIENT SAMPLE RESULTS WERE NOT AFFECTED AND WOULD NOT BE AFFECTED IF THE EVENT WERE TO RECUR UNDETECTED. THERE WERE NO ALLEGATIONS OF PATIENT HARM. THIS REPORT IS NUMBER NINE OF TWENTY MDR'S FOR THIS EVENT. TWENTY 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWENTY DEVICES WERE INVOLVED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 760612 | VITROS CHEMISTRY PRODUCTS LAC SLIDES | IN VITRO DIAGNOSTIC | KHP | ORTHO-CLINICAL DIAGNOSTICS | 3530-0092-3271 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |