Report
- Report Number
- 2134265-2016-10161
- Event Type
- Malfunction
- Date Received
- November 17, 2016
- Date of Event
- October 22, 2016
- Report Date
- October 22, 2016
- Manufacturer
- BOSTON SCIENTIFIC - MAPLE GROVE
- Product Code
- DXE
- PMA / PMN Number
- K151313
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. (B)(4). DEVICE RETURNED TO MFR.:RETURNED PRODUCT CONSISTED OF A ZELANTEDVT CATHETER SYSTEM AND NO OTHER DEVICES. THE PUMP, SHAFT, AND TIP WERE MICROSCOPICALLY EXAMINED AND NO DAMAGE OR IRREGULARITIES WERE IDENTIFIED. FUNCTIONAL TESTING WAS PERFORMED BY PLACING THE DEVICE IN THE CONSOLE. FUNCTIONAL TESTING SHOWED A LEAK AT THE MALE CONNECTOR WITH FEMALE LUER. FLUID LEAKING FROM THE CONNECTOR IS IN INDICATION THAT THE OUTLET ADAPTER SEAL FAILED. INSPECTION OF THE REMAINDER OF THE DEVICE, APART FROM THE OBSERVED DAMAGE, REVEALED NO OTHER DAMAGE OR IRREGULARITIES. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE OF THE REPORTED DIFFICULTY IS A DESIGN CONSTRAINT OF THE PRODUCT. (B)(4).
SAME CASE AS: 2134265-2016-10160. IT WAS REPORTED THAT AN ERROR MESSAGE DISPLAYED DURING ASPIRATION. AN ANGIOJET® ZELANTEDVT¿ WAS SELECTED FOR A THROMBECTOMY PROCEDURE IN THE ILIOFEMORAL VEIN. THE CATHETER WAS SUCCESSFULLY PRIMED AND ASPIRATION WAS INITIATED. AFTER ABOUT 90 SECONDS, THE DEVICE QUIT WORKING AND AN ERROR MESSAGE TO "CHECK SALINE SUPPLY" DISPLAYED ON THE CONSOLE. THEY ATTEMPTED TO RESOLVE THE ISSUE BY RESETTING THE SYSTEM. THE DEVICE WOULD RETURN TO FUNCTION FOR 2-3 SECONDS, BUT THEN STOP AGAIN WITH THE SAME ERROR MESSAGE. ANOTHER ANGIOJET® ZELANTEDVT¿ WAS SELECTED FOR USE. HOWEVER, THE SAME ISSUE OCCURRED. THE CATHETER WAS SUCCESSFULLY PRIMED, BUT FAILED WITH THE ERROR MESSAGE AFTER 5-10 SECONDS OF USE INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH BALLOONS AND STENTS. THERE WERE NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT CONDITION WAS NOTED AS STABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 759775 | ANGIOJET® ZELANTEDVT¿ | CATHETER, EMBOLECTOMY | DXE | BOSTON SCIENTIFIC - MAPLE GROVE | 114610-001 | 19776370 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |