FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 6109490 · Received November 17, 2016

Report

Report Number
1045834-2016-13243
Event Type
Malfunction
Date Received
November 17, 2016
Report Date
September 8, 2016
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: (B)(4). (B)(6) THE DEVICE WAS RETURNED FOR SERVICE; HOWEVER, DID NOT MEET MANUFACTURING SPECIFICATIONS DURING PRE-REPAIR ASSESSMENT. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO IMPROPER HANDLING WHICH IS USER ERROR/MISUSE/ABUSE. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED BY (B)(6) THAT DURING SERVICE AND EVALUATION, IT WAS OBSERVED THAT THE CABLE/CORD/WIRING WAS DAMAGED ON THE MOTOR DEVICE. IT WAS FURTHER OBSERVED THAT THE MOTOR AND CONTROL WERE DEFECTIVE, AN ERROR E6 WAS DISPLAYED, THE COUPLING WAS WORN OUT, THE DEVICE WAS GETTING HOT, AND THE COVER WAS MISSING. IT WAS ALSO NOTED THAT THE DEVICE FAILED PRE-TEST FOR LOCTITE AND CABLE ASSESSMENTS AND SAFETY. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762362 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1