FDA Adverse Event Malfunction Summary report: N

11CM ANGLE ATTACHMENT

MDR report key: 6109482 · Received November 17, 2016

Report

Report Number
1045834-2016-13248
Event Type
Malfunction
Date Received
November 17, 2016
Date of Event
November 1, 2016
Report Date
November 2, 2016
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
K011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE FAILED THE TEMPERATURE TEST DUE TO WORN OUT/BROKEN BEARINGS. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WORN OUT BEARINGS FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: (B)(4). THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE RELIABILITY ENGINEERING EVALUATES THE DEVICE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE WAS OVERHEATING. THERE WERE NO DELAYS TO THE SURGICAL PROCEDURE. IT WAS NOT REPORTED IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS REPORTED THAT THE SURGERY WAS SUCCESSFULLY COMPLETED WITH NO FURTHER INCIDENCE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
761505 11CM ANGLE ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS N/A

Patients

Seq Age Sex Outcome Treatment
1