FDA Adverse Event Malfunction Summary report: N

530G INSULIN PUMP

MDR report key: 6107781 · Received November 16, 2016

Report

Report Number
2032227-2016-42389
Event Type
Malfunction
Date Received
November 16, 2016
Date of Event
November 10, 2016
Report Date
December 12, 2016
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4).

Additional Manufacturer Narrative · 1

THE PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO AN UNLOCKED CONNECTOR ON THE LCD BOARD. NO AUDIO/BEEP ANOMALY OR FROZEN DISPLAY WAS NOTED. UNABLE TO PERFORM ANY TESTS DUE TO UNRESPONSIVE BUTTONS. THE PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP AND MINOR SCRATCHES ON THE DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE CALL TO HAVE EXPERIENCED A LOUD CONSTANT TONE OF INSULIN PUMP. CUSTOMER CHANGED BATTERIES AND CONSTANT TONE WAS NOT RESOLVED. TROUBLESHOOTING COULD NOT BE PERFORMED AS INSULIN PUMP ALSO HAD A BLANK DISPLAY SCREEN. CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. CUSTOMER'S BLOOD GLUCOSE WAS 112 MG/DL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
758225 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAS

Patients

Seq Age Sex Outcome Treatment
1 45 YR