FDA Adverse Event Malfunction Summary report: N

PUMP IN STYLE ADVANCED - ON-THE-GO TOTE

MDR report key: 6107177 · Received November 16, 2016

Report

Report Number
1419937-2016-00296
Event Type
Malfunction
Date Received
November 16, 2016
Date of Event
October 25, 2016
Report Date
November 16, 2016
Manufacturer
MEDELA, INC
Product Code
HGX
PMA / PMN Number
K031614
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REPLACEMENT POWER SUPPLY WAS SENT TO THE CUSTOMER. THE ORIGINAL PRODUCT WAS RECEIVED ON 11/16/2016. HOWEVER, A PRODUCT EVALUATION HAD NOT BEEN COMPLETED AT THE TIME OF THIS REPORT. THIS ISSUE WITH THE DAMAGED TRANSFORMER HOUSING FOR THE PUMP IN STYLE DEVICE WAS ADDRESSED IN INVESTIGATION (B)(4). THE INVESTIGATION FOUND THAT THE TRANSFORMERS ARE BEING DAMAGED DURING SHIPMENT FROM THE MANUFACTURER TO MEDELA. THIS DAMAGE IS CAUSING THE PLASTIC HOUSING TO FAIL PREMATURELY WHEN SUBJECTED TO NORMAL USE AND FORESEEABLE MISUSE. THE PACKAGING USED BY THE MANUFACTURER TO SHIP THE TRANSFORMERS TO MEDELA IS NOT ROBUST ENOUGH TO HANDLE ALL OF THE POTENTIAL SHIPPING, HANDLING, AND ABUSE CONDITIONS THAT COULD ARISE FROM LOGISTICS OF THE CONSOLIDATION PROCESS. AS A RESULT OF THE INVESTIGATION, THE SHIPPING AND CONSOLIDATION PROCESS HAS BEEN MODIFIED TO REDUCE THE HANDLING AND POTENTIAL FOR DOUBLE STACKING OF THE SKIDS. THE SHIPPING PACKAGING STRENGTH HAS ALSO BEEN INCREASED TO FURTHER PROTECT THE TRANSFORMERS DURING SHIPPING. COMPLAINTS AGAINST THIS PRODUCT ARE CURRENTLY BEING MONITORED FOR EFFECTIVENESS OF THE ABOVE MENTIONED CORRECTIVE ACTION.

Description of Event or Problem · 1

ON (B)(6) 2016 THE CUSTOMER REPORTED TO A MEDELA CUSTOMER SERVICE REPRESENTATIVE THAT THE POWER ADAPTER FOR HER PUMP IN STYLE BREAST PUMP CAME APART.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
758436 PUMP IN STYLE ADVANCED - ON-THE-GO TOTE PUMP, BREAST, POWERED HGX MEDELA, INC 57063 / 9207010 5112 / REV M

Patients

Seq Age Sex Outcome Treatment
1