O-ARM 1000 IMAGING SYSTEM
Report
- Report Number
- 1723170-2016-04017
- Event Type
- Malfunction
- Date Received
- November 16, 2016
- Date of Event
- November 19, 2014
- Report Date
- November 16, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE REPRESENTATIVE FOUND THAT THE NUMBER OF PATIENTS ON THE IMAGING SYSTEM EXCEEDED 300. AFTER CLEARING OUT THE PATIENT DATA; THE HARDWARE, SOFTWARE, AND INSTRUMENTS PASSED THE SYSTEM CHECKOUT. THE SYSTEM WAS FOUND TO BE FULLY FUNCTIONAL. THE SOFTWARE INVESTIGATION FOUND THAT THE REPORTED EVENT WAS RELATED TO A SOFTWARE ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION SOFTWARE ANOMALY TRACKING DATABASE. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(6) ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 FDA-483 FEI: 3004785967. THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
A MEDTRONIC REPRESENTATIVE REPORTED THAT WHEN STARTING UP THE VIEWING STATION OF THE IMAGING SYSTEM, AN ERROR MESSAGE APPEARED STATING "AN ERROR HAS OCCURRED THAT MAY HAVE DAMAGED THE SYSTEMS DATABASE, PLEASE CONTACT MEDTRONIC TECHNICAL SERVICES." INITIAL TROUBLESHOOTING PERFORMED INCLUDED RESTARTING THE SYSTEM, THEN CREATING A PATIENT IMAGE BACKUP THROUGH PHONE-CALL WITH TECHNICAL SERVICES (TS). THE BACKUP GENERATION RESOLVED THE REPORTED ISSUE. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 755946 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |