FDA Adverse Event Injury Summary report: N

T:SLIM G4 SYSTEM

MDR report key: 6105837 · Received November 16, 2016

Report

Report Number
3007981285-2016-19299
Event Type
Injury
Date Received
November 16, 2016
Date of Event
October 26, 2016
Report Date
October 26, 2016
Manufacturer
TANDEM DIABETES CARE
Product Code
OYC
PMA / PMN Number
P140015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION; HOWEVER, DEVICE EVALUATION IS NOT YET COMPLETE. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PUMP COULD NOT CHARGE PROPERLY WITHOUT THE CUSTOMER MANEUVERING THE CABLE INTO THE CHARGING PORT AT A CERTAIN ANGLE, IN ORDER TO COMPLETE A SUCCESSFUL CHARGE. THE CUSTOMER ATTEMPTED MULTIPLE CABLES AND STATED THAT THE USB CABLES FIT VERY LOOSELY INTO THE USB PORT. CUSTOMER REPORTED BLOOD GLUCOSE (BG) LEVEL WAS IMPACTED (336 MG/DL). THE CUSTOMER STATED THEY HAD INSULIN ON BOARD TO ADDRESS BG LEVEL. REPORTEDLY THE CUSTOMER WILL CONTINUE TO USE THE PUMP UNTIL THE REPLACEMENT PUMP IS RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
757588 T:SLIM G4 SYSTEM INSULIN PUMP OYC TANDEM DIABETES CARE 4628-003

Patients

Seq Age Sex Outcome Treatment
1 27 YR Other