FDA Adverse Event Other Summary report: N

ARJO AMBULIFT BSQ

MDR report key: 61050 · Received January 3, 1997

Report

Report Number
1922538-1997-00003
Event Type
Other
Date Received
January 3, 1997
Date of Event
December 16, 1996
Report Date
January 2, 1997
Manufacturer
ARJO MANUFACTURING CO.
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE FOLLOWING IS A CLOSE OUT REPORT BY THE FOREIGN REPORTER. INVESTIGATION: A WHEEL ON THE SUB-CHASSIS WAS ALSO FOUND TO BE LOOSE AND HAD FELL OUT, THIS WAS RE-TIGHTENED. THE BACK REST SCREWS WERE LOOSE AND TWO WERE MISSING ALTOGETHER. OBSERVATIONS: 5 OTHER DEVICES

Description of Event or Problem · 1

THE PICK UP HOOK OF THE JIB HAS BECOME DEFORMED AND ALLOWED THE CHAIR TO BE DISLODGED AS A PT WAS BEING HOISTED INTO THE BATH. THE CHAIR AND PT DID NOT FALL TO THE FLOOR AND INJURY HAS NOT OCCURRED. ONE PERSON INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARJO AMBULIFT BSQ PATIENT HOIST FSA ARJO MANUFACTURING CO. AB1402 NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other