FDA Adverse Event Malfunction Summary report: N

O-ARM 1000 IMAGING SYSTEM

MDR report key: 6104786 · Received November 15, 2016

Report

Report Number
1723170-2016-03342
Event Type
Malfunction
Date Received
November 15, 2016
Date of Event
June 24, 2015
Report Date
November 15, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LITTLETON)
Product Code
OXO
PMA / PMN Number
K050996
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE SYSTEM. REPLACEMENT MOBILE VIEW STATION (MVS) POWER FUSES WERE SHIPPED TO THE SITE. AFTER REPLACING THE POWER FUSES, THE MEDTRONIC REPRESENTATIVE PERFORMED A SYSTEM CHECKOUT . THE HARDWARE, SOFTWARE, AND INSTRUMENTS PASSED THE SYSTEM CHECKOUT. THE SYSTEM WAS FOUND TO BE FULLY FUNCTIONAL. THIS PART WAS DISCARDED BY THE SITE AND WILL NOT BE RETURNED TO MANUFACTURER FOR ANALYSIS. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA (B)(6) DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.

Description of Event or Problem · 1

A SITE REPRESENTATIVE REPORTED THAT THE IMAGE ACQUISITION SYSTEM (IAS) WAS NOT POWERING ON OR CHARGING. THE ORANGE POWER LIGHT WAS LIT ON THE IAS. THE SITE REQUESTED THE MEDTRONIC FIELD SERVICE ENGINEER PERFORM A SYSTEM CHECKOUT ON THE IAS. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
754454 O-ARM 1000 IMAGING SYSTEM IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE OXO MEDTRONIC NAVIGATION, INC. (LITTLETON) BI-700-00027-120

Patients

Seq Age Sex Outcome Treatment
1