FDA Adverse Event Malfunction Summary report: N

O-ARM 1000 IMAGING SYSTEM 3RD EDITION

MDR report key: 6104643 · Received November 15, 2016

Report

Report Number
1723170-2016-05233
Event Type
Malfunction
Date Received
November 15, 2016
Date of Event
February 1, 2016
Report Date
November 15, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LITTLETON)
Product Code
OXO
PMA / PMN Number
K092564
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
RADIOLOGIC TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

PATIENT WEIGHT NOT AVAILABLE FROM THE SITE. A SITE REPRESENTATIVE REPORTED THAT, WHILE IN A SPINAL FUSION PROCEDURE, THEY WERE UNABLE TO TAKE A 3D SPIN. FOLLOWING A SYSTEM REBOOT THEY LOST THE ABILITY TO TAKE 2D IMAGES AS WELL AS 3D IMAGES. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME. ONSITE INVESTIGATION WAS PERFORMED AND THE FIELD SYSTEMS ENGINEER WAS UNABLE TO REPLICATE THE ISSUE, HOWEVER, HE NOTED THAT IT IS A KNOWN SOFTWARE PROBLEM. NO FURTHER ISSUES WERE REPORTED. A FULL IMAGING SYSTEM CHECK-OUT WAS COMPLETED AND ALL TESTS PASSED. FULL SYSTEM FUNCTIONALITY WAS CONFIRMED AND THE SYSTEM WAS RETURNED TO SERVICE. (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.

Description of Event or Problem · 1

A SITE REPRESENTATIVE REPORTED THAT, WHILE IN A SPINAL FUSION PROCEDURE, THEY WERE UNABLE TO TAKE A 3D SPIN. FOLLOWING A SYSTEM REBOOT THEY LOST THE ABILITY TO TAKE 2D IMAGES AS WELL AS 3D IMAGES. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
752333 O-ARM 1000 IMAGING SYSTEM 3RD EDITION IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE OXO MEDTRONIC NAVIGATION, INC. (LITTLETON) BI-700-00028-120

Patients

Seq Age Sex Outcome Treatment
1 55 YR