FDA Adverse Event Malfunction Summary report: N

PRISM CHAGAS

MDR report key: 6103038 · Received November 15, 2016

Report

Report Number
1415939-2016-00112
Event Type
Malfunction
Date Received
November 15, 2016
Date of Event
July 10, 2016
Report Date
December 21, 2016
Manufacturer
ABBOTT LABORATORIES
Product Code
MIU
UDI-DI
00380740062941
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. AN EVALUATION IS IN-PROCESS.

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, IN-HOUSE TESTING, A DEVICE HISTORY REVIEW, A SEARCH FOR SIMILAR COMPLAINTS, AND A REVIEW OF LABELING. RETURN MATERIAL WAS NOT AVAILABLE FROM THE CUSTOMER. A SPECIFICITY TESTING PROTOCOL WAS EXECUTED USING LOT 60017M500 AND MET ACCEPTANCE CRITERIA AND DETERMINED THE REAGENT IS PERFORMING ACCEPTABLY. THE DEVICE HISTORY REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES OR DEVIATIONS RELATED TO THE CUSTOMER'S OBSERVATIONS. TRACKING AND TRENDING DID NOT IDENTIFY AN ADVERSE TREND. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE AVAILABLE INFORMATION NO PRODUCT DEFICIENCY OF THE PRISM CHAGAS REAGENT, LIST NUMBER (B)(4), LOT NUMBER 60017M500, WAS IDENTIFIED.

Description of Event or Problem · 1

THE CUSTOMER INDICATED AN INCREASE IN REACTIVE RESULTS HAS BEEN GENERATED WHILE USING THE PRISM CHAGAS ASSAY ON THE PRISM 6 CHANNEL ANALYZER. DURING THE TIMEFRAME OF (B)(6) 2016 TO (B)(6) 2016, TWELVE SPECIMENS WHICH HAD BEEN REPEAT REACTIVE WERE NEGATIVE WHEN TESTED USING THE ORTHO METHOD. SPECIFICALLY THE CUSTOMER INDICATED THAT FALSE POSITIVE RESULTS WERE GENERATED ON (B)(6) 2016 AND (B)(6) 2016 USING LOT 60017M500. NO SPECIFIC PATIENT DATA HAS BEEN PROVIDED. THERE WAS NO ADVERSE IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
753964 PRISM CHAGAS CHAGAS MIU ABBOTT LABORATORIES 60017M500 00380740062941

Patients

Seq Age Sex Outcome Treatment
1 LN (B)(4), SN (B)(4)| PRISM 6 CHANNEL ANALYZER