FDA Adverse Event Malfunction Summary report: N

DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 6102979 · Received November 15, 2016

Report

Report Number
3004753838-2016-84992
Event Type
Malfunction
Date Received
November 15, 2016
Date of Event
October 24, 2016
Report Date
October 24, 2016
Manufacturer
DEXCOM, INC.
Product Code
MDS
UDI-DI
00386270000279
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4)

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016, THAT ON (B)(6) 2016, THE RECEIVER DISPLAYED ERR121. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE. THE RECEIVER WAS RETURNED FOR EVALUATION. AN EXTERNAL VISUAL INSPECTION WAS PERFORMED AND PASSED. THE RECEIVER DATA LOG WAS DOWNLOADED AND REVIEWED AND FIRMWARE ERRORS WERE OBSERVED. THE REPORTED EVENT OF THE RECEIVER DISPLAYING ERR121 WAS CONFIRMED. A ROOT CAUSE COULD NOT BE DETERMINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
755048 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM, INC. MT22719-BLU 5205056 00386270000279

Patients

Seq Age Sex Outcome Treatment
1 20 YR