FDA Adverse Event Malfunction Summary report: N

MEDLINE RENEWAL REPROCESSED SCORPION NEEDLE

MDR report key: 6102890 · Received November 15, 2016

Report

Report Number
3032391-2016-00042
Event Type
Malfunction
Date Received
November 15, 2016
Date of Event
October 13, 2016
Report Date
November 15, 2016
Manufacturer
MEDLINE RENEWAL
Product Code
HWQ
PMA / PMN Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE IN QUESTION WAS RETURNED TO MEDLINE RENEWAL FOR EVALUATION. UPON RECEIPT, MEDLINE RENEWAL CONFIRMED THE DEVICE HAD BEEN REPROCESSED. OUR EVALUATION CONFIRMED THE REPORT, HOWEVER THE TIP WAS NOT RETURNED FOR EVALUATION. OUR INVESTIGATION INCLUDED BOTH AN INSPECTION OF THE RETURNED DEVICE, AND A REVIEW OF THE DEVICE HISTORY RECORD (DHR). WE RECONFIRMED THAT ALL PROCESSES WERE CONDUCTED AS REQUIRED, AND THAT THE DEVICE MET INSPECTION AND REQUIREMENTS PRIOR TO PACKAGING AND RELEASE. UNFORTUNATELY, THE ROOT CAUSE OF THE DEVICE FAILURE IS UNKNOWN AT THIS TIME. PER THE DEVICE INSTRUCTIONS FOR USE (IFU), THE USE OF EXCESSIVE FORCE TO PASS THE NEEDLE THROUGH FIBROUS/CALCIFIC TISSUE, OR STRIKING BONE CAN CAUSE NEEDLE BREAKAGE AND PATIENT INJURY. THIS CAN BE AVOIDED BY REPOSITIONING THE NEEDLE PASS THROUGH A DIFFERENT AREA. THERE HAS BEEN NO REPORT OF ADVERSE PATIENT CONSEQUENCE AS A RESULT OF THE INCIDENT. HOWEVER, DUE TO THE SURGICAL INTERVENTION TO RETRIEVE THE TIP WHICH CAUSED A DELAY IN THE PROCEDURE, MEDLINE RENEWAL IS FILING THIS MEDWATCH REPORT.

Description of Event or Problem · 1

MEDLINE RENEWAL RECEIVED A REPORT INDICATING THAT THE TIP OF A REPROCESSED ARTHREX SUREFIRE SCORPION NEEDLE, MODEL AR-13991N, DETACHED DURING USE. THE TIP WAS RETRIEVED BY THE SURGEON. ANOTHER DEVICE WAS READILY AVAILABLE TO COMPLETE THE PROCEDURE; THEREFORE NO PATIENT HARM OR ADDITIONAL MEDICAL INTERVENTION WAS REPORTED AS A RESULT OF THE INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
753871 MEDLINE RENEWAL REPROCESSED SCORPION NEEDLE SUTURE PASSER HWQ MEDLINE RENEWAL AR-13991N 312381

Patients

Seq Age Sex Outcome Treatment
1 44 YR Required Intervention