FDA Adverse Event Injury Summary report: N

ACHIEVA 1.5T NOVA

MDR report key: 6102109 · Received November 15, 2016

Report

Report Number
3003768277-2016-00106
Event Type
Injury
Date Received
November 15, 2016
Date of Event
October 18, 2016
Report Date
October 18, 2016
Manufacturer
PHILIPS HEALTHCARE
Product Code
LNH
PMA / PMN Number
K052013
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE PROVIDED INFORMATION AND INVESTIGATION PERFORMED NO MALFUNCTION OF THE MR DEVICE WAS OBSERVED. THE FINGER SWITCH PLATE WAS TESTED ON SITE, AND FOUND TO BE WORKING. IT WAS HOWEVER REPORTED THAT IN THIS CASE NO ARM SUPPORTS WERE USED TO ENSURE THAT THE PATIENT'S EXTREMITIES REMAINED ON THE TABLETOP. ARM SUPPORTS ARE DELIVERED WITH EACH MRI SYSTEM. THESE ARM SUPPORTS CAN BE USED TO AVOID FINGER PINCHING.

Description of Event or Problem · 1

PHILIPS RECEIVED A REPORT FROM A CUSTOMER THAT A PATIENT'S FINGER GOT INJURED WHILE BEING MOVED INTO THE BORE OF THE MR SYSTEM. THE FINGER OF THE PATIENT WAS PINCHED BETWEEN THE TABLE TOP AND THE MAGNET BORE COVER, RESULTING INTO A FRACTURE OF THE FINGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
752515 ACHIEVA 1.5T NOVA SYSTEM, NUCLEAR MAGNETIC RESONANCE IMAGING LNH PHILIPS HEALTHCARE 781196

Patients

Seq Age Sex Outcome Treatment
1 Other