FDA Adverse Event Malfunction Summary report: N

PALMTOP VENTILATOR, MODELS PTV-8, PTV-10

MDR report key: 6100981 · Received November 14, 2016

Report

Report Number
2031702-2016-01403
Event Type
Malfunction
Date Received
November 14, 2016
Date of Event
April 29, 2014
Report Date
November 14, 2016
Manufacturer
CAREFUSION, INC
Product Code
CBK
PMA / PMN Number
K070594
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE AC ADAPTER WAS RECEIVED FOR EVALUATION. AN EVALUATION OF THE DEVICE VERIFIED THE REPORTED ISSUE AND FOUND THE AC LEMO CONNECTOR PIN #1 IS BURNT/MELTED AND THE AC ADAPTER CABLE CAME APART FROM THE LEMO CONNECTOR, EXPOSING AWG WIRES. THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE. THIS COMPLAINT WAS INCLUDED IN A TWO-YEAR RETROSPECTIVE COMPLAINT REVIEW TO ASSESS MDR REPORTABILITY COMPLIANCE. THIS COMPLAINT WAS DEEMED TO MEET THE REQUIREMENTS FOR MDR SUBMISSION AND IS THEREFORE BEING SUBMITTED TO THE FDA.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEMO CONNECTOR HAS FALLEN APART. WHILE IN SERVICE, IT WAS DISCOVERED THAT THE UNIT HAD BURNT COMPONENTS. THIS REPORTABLE ISSUE WAS DISCOVERED WHILE IN SERVICE, THEREFORE, THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
750764 PALMTOP VENTILATOR, MODELS PTV-8, PTV-10 VENTILATOR, CONTINUOUS, FACILITY USE CBK CAREFUSION, INC PTV AC ADAPTER

Patients

Seq Age Sex Outcome Treatment
1