PASSPORT 17 M PATIENT MONITOR
Report
- Report Number
- 3007222337-2016-00033
- Event Type
- Malfunction
- Date Received
- November 14, 2016
- Date of Event
- November 1, 2016
- Report Date
- December 22, 2016
- Manufacturer
- SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
- Product Code
- MHX
- PMA / PMN Number
- K143195
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
FIVE DEFECTIVE HOSES WERE RETURNED FOR EVALUATION. IT WAS DETERMINED THAT THE LEAK WAS CAUSED BY THE DETERIORATION OF THE INNER O-RING OF THE NIBP HOSE, DUE TO THE USE OF NON-RECOMMENDED CLEANING AGENT BY THE CUSTOMER. THE CUSTOMER WAS PROVIDED A LIST OF RECOMMENDED CLEANING AGENTS, ALSO LOCATED IN THE OPERATOR'S MANUAL. MINDRAY DS, USA, INC. ((B)(4)). (EXEMPTION # E2015032).
(B)(4). (EXEMPTION # E2015032).
IT WAS REPORTED THAT THE NON INVASIVE BLOOD PRESSURE HOSE USED WITH THE PASSPORT 12M PATIENT MONITOR WAS LEAKING AND THE MONITOR PRESENTED AN ERRONEOUS VALUE. THE HOSPITAL STAFF TREATED THE PATIENT UNDER THE IMPRESSION THAT THESE WERE ACCURATE READINGS AND ADMINISTERED MEDICATION. IT WAS ONLY AFTER THE PATIENT DID NOT RESPOND TO THE MEDICATION THAT THE BLOOD PRESSURE READINGS WERE FOUND TO BE REPORTING LOWER THAN ACTUAL.
IT WAS REPORTED THAT THE NON INVASIVE BLOOD PRESSURE HOSE USED WITH THE PASSPORT 12M PATIENT MONITOR WAS LEAKING AND THE MONITOR PRESENTED AN ERRONEOUS VALUE. THE HOSPITAL STAFF TREATED THE PATIENT UNDER THE IMPRESSION THAT THESE WERE ACCURATE READINGS AND ADMINISTERED MEDICATION. IT WAS ONLY AFTER THE PATIENT DID NOT RESPOND TO THE MEDICATION THAT THE BLOOD PRESSURE READINGS WERE FOUND TO BE REPORTING LOWER THAN ACTUAL.