FDA Adverse Event Malfunction Summary report: N

DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 6100171 · Received November 14, 2016

Report

Report Number
3004753838-2016-28441
Event Type
Malfunction
Date Received
November 14, 2016
Date of Event
October 11, 2016
Report Date
October 20, 2016
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Removal / Correction Number
3004753838-02/29/16-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE RECEIVER WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. A MANUAL TEST WAS PERFORMED AND SPEAKER DID NOT SOUND. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.

Description of Event or Problem · 1

DEXCOM WAS MADE AWARE ON (B)(6) 2016, TO CLAIM NO AUDIO OUTPUT ON (B)(6) 2016. THERE WAS NO REPORT OF ANY INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
751982 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM CONTINUOUS GLUCOSE MONITOR MDS DEXCOM, INC. MT20649-2 5210473

Patients

Seq Age Sex Outcome Treatment
1