ALARIS SECONDARY SET
Report
- Report Number
- 9616066-2016-01562
- Event Type
- Malfunction
- Date Received
- November 14, 2016
- Date of Event
- October 13, 2016
- Report Date
- October 26, 2016
- Manufacturer
- CAREFUSION
- Product Code
- FPA
- PMA / PMN Number
- K053049
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHARMACIST
Narratives
CONCOMITANT MEDICAL PRODUCTS: 24301-0007T; B.BRAUN 1000ML BAG OF 0.9% NACL INJECTION, NDC 0264-7800-09, EXP DATE 03/18, LOT # J5L763; B.BRAUN 250ML DOCETAXEL (TAXOTERE), NDC 0264-7800-20, EXP DATE 06/18, LOT # J6H262; B.BRAUN 250ML RAMUCIRUMAB, NDC 0264-7800-20, EXP DATE 06/18, LOT # J6H262, THERAPY DATE (B)(6) 2016. THE AFFECTED PRODUCT HAS BEEN RECEIVED AND THE INVESTIGATION IS PENDING. A FOLLOW UP REPORT WILL BE SUBMITTED ONCE THE FAILURE INVESTIGATION HAS BEEN COMPLETED.
THE CUSTOMER¿S REPORT OF ¿LEAK¿ WAS NOT CONFIRMED. THE SETS WERE VISUALLY INSPECTED FOR KINKS, HOLES/TEARS IN THE TUBING OR DAMAGES TO THE COMPONENTS. NO ANOMALIES WERE OBSERVED ON THE SET. FUNCTIONAL AND PRESSURE TESTING RESULTED IN NO SIGNS OF LEAKING OR OCCLUSION ANYWHERE ON THE RETURNED SET WHILE THE TUBING WAS MANIPULATED AT EACH ENGAGEMENT. THE ROOT CAUSE OF THE CUSTOMER¿S REPORT OF A ¿LEAK¿ WAS NOT IDENTIFIED.
THE CUSTOMER REPORTED THAT A PATIENT WAS RECEIVING TAXOTERE CHEMOTHERAPY. WHILE ADDRESSING THE END OF INFUSION ALARM AND SWITCHING TO THE PRIMARY SALINE INFUSION, THE NURSE NOTED LIQUID ON HIS/HER THUMB AND A FEW DROPS ON THE BOTTOM OF THE IV POLE, DETERMINED TO BE LEAKING FROM THE BONDED, CLOSED SYSTEM VALVE ATTACHED TO THE CHEMO SECONDARY TUBING. IT WAS DETERMINED THAT LESS THAN 5ML HAD LEAKED. THE NURSE FLUSHED AND WASHED HANDS FOR 6-8 MINUTES, AND NO HARM TO THE NURSE OR PATIENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 750975 | ALARIS SECONDARY SET | SET, SECONDARY, INTRAVASCULAR | FPA | CAREFUSION | 10013364T |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | 11171447,8100,8000,10012241-0500 |