O-ARM 1000 IMAGING SYSTEM
Report
- Report Number
- 1723170-2016-04526
- Event Type
- Malfunction
- Date Received
- November 14, 2016
- Date of Event
- December 23, 2014
- Report Date
- November 14, 2016
- Manufacturer
- MEDTRONIC NAVIGATION, INC. (LITTLETON)
- Product Code
- OXO
- PMA / PMN Number
- K050996
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
PATIENT INFORMATION WAS UNAVAILABLE FROM THE SITE. A MEDTRONIC FIELD SERVICE ENGINEER (FSE) WENT TO SITE TO TROUBLESHOOT ISSUE. IT WAS FOUND THERE WERE FRAME RATE ERRORS DURING AN IMAGING SPIN ON 2 OCCASIONS. ( THIS IS THE 2ND OCCURRENCE) AFTER TESTING THE FSE COULD NOT DUPLICATE FRAME RATE ERRORS. LVDS CABLE CHECKED. DECISION MADE TO REPLACE UMBILICAL AS PRECAUTION. UMBILICAL CABLE SENT TO SITE AND SWAPPED OUT. SYSTEM CHECKED OUT AFTER REPLACEMENT AND PASSED ALL TESTING AND PUT BACK INTO USE. THE CABLE WAS SENT BACK TO MANUFACTURER FOR EVALUATION. CONFIRMED REPORTED PROBLEM. UMBILICAL FAILED CONTINUITY TEST. J8 PIN 2 (BROWN WIRE ) WAS CUT. NO FURTHER ISSUES REPORTED. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA NEW ENGLAND DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO (B)(4). THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.
A FIELD SERVICE ENGINEER (FSE) REPORTED A FRAME RATE ERROR MESSAGE ON THE IMAGING SYSTEM. ISSUE WAS NOTICED WHILE TAKING AN IMAGING SPIN DURING A SURGERY. THIS IS THE SECOND OCCURRENCE. FIRST OCCURRENCE IS IN (B)(4). THERE WAS NO IMPACT ON PATIENT OUTCOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 750219 | O-ARM 1000 IMAGING SYSTEM | IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE | OXO | MEDTRONIC NAVIGATION, INC. (LITTLETON) | BI-700-00027-120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 36 YR |