FDA Adverse Event
Malfunction
Summary report: N
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 6099273
·
Received November 14, 2016
Report
- Report Number
- 3004753838-2016-54140
- Event Type
- Malfunction
- Date Received
- November 14, 2016
- Date of Event
- October 17, 2016
- Report Date
- October 17, 2016
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Removal / Correction Number
- 3004753838-02/29/16-001C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4)
Additional Manufacturer Narrative · 1
(B)(4)
Description of Event or Problem · 1
THE RECEIVER WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECTS WERE FOUND. A MANUAL TEST WAS PERFORMED AND SPEAKER DID NOT SOUND. THE REPORTED EVENT OF NO AUDIO OUTPUT WAS CONFIRMED. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.
Description of Event or Problem · 1
DEXCOM WAS MADE AWARE ON (B)(6) 2016, TO CLAIM NO AUDIO OUTPUT ON (B)(6) 2016. THERE WAS NO REPORT OF ANY INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL EVENT OR PATIENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 750894 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | MDS | DEXCOM, INC. | MT20649-2 | 5210527 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |