O-ARM 1000 IMAGING SYSTEM 3RD EDITION

MDR report key: 6097208 · Received November 11, 2016

Report

Report Number
1723170-2016-05028
Event Type
Malfunction
Date Received
November 11, 2016
Date of Event
July 30, 2015
Report Date
November 11, 2016
Manufacturer
MEDTRONIC NAVIGATION, INC. (LITTLETON)
Product Code
OXO
PMA / PMN Number
K092564
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

NO PATIENT INFORMATION PROVIDED AS NO PATIENT WAS INVOLVED IN THIS CONCERN. A MEDTRONIC REPRESENTATIVE REPLACED THE IMAGING SYSTEM'S MOTION BATTERIES AND THE TESTED THE EQUIPMENT. THE HARDWARE, SOFTWARE, AND INSTRUMENTS PASSED THE SYSTEM CHECKOUT. NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. THIS EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW AS DISCUSSED WITH THE FDA NEW ENGLAND DISTRICT OFFICE ON APRIL 7, 2016 VIA MEDTRONIC NAVIGATION, INC. RESPONSE TO 3/17/2016 FDA-483 FEI: 3004785967. THIS REVIEW WAS PERFORMED AS A RESULT OF RECENT CHANGES/IMPROVEMENTS TO THE MEDTRONIC NAVIGATION, INC. MEDICAL DEVICE REPORT (MDR) REVIEW PROCESS AND IS FOR NON-ADVERSE EVENT REPORTABLE MALFUNCTIONS WITHIN THE LAST TWO YEARS WHERE THE REPORTED MALFUNCTION WAS NOT PREVIOUSLY ASSOCIATED WITH SERIOUS INJURIES. SIMILAR MALFUNCTIONS THAT HAD RESULTED IN SIGNIFICANT DELAYS OF SURGERY OR REQUIRED ADDITIONAL INTERVENTION HAD BEEN REPORTED AS MDR SERIOUS INJURIES. BASED ON THIS 483 OBSERVATION, THE REPORTING DETERMINATION HAS BEEN REVISED TO CONSIDER THESE MALFUNCTION EVENTS REPORTABLE. THIS PROCESS CHANGE RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY MEDTRONIC NAVIGATION, INC., SINCE APRIL 2016. IT SHOULD BE NOTED THAT THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE REGULATIONS WITHIN OUR PROCESSES.

Description of Event or Problem · 1

A MEDTRONIC REPRESENTATIVE REPORTED THAT THE MEDTRONIC IMAGING SYSTEM'S MOTION BATTERY CHARGE DROPS RAPIDLY WHEN THE UMBILICAL CORD IS DISCONNECTED AND THE UNIT IS THEN DRIVEN TO THE OPERATING ROOM. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.