FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 6095012 · Received November 10, 2016

Report

Report Number
2531779-2016-31177
Event Type
Malfunction
Date Received
November 10, 2016
Report Date
October 18, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION: 12/29/2016. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/08/2016 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE BATTERY COMPARTMENT WAS CRACKED ON THE SIDE FROM THE THREADS TO THE COVER. UNRELATED TO THE ORIGINAL COMPLAINT, THE DISPLAY SCREEN HAS A PINKISH CONTRAST. ADDITIONALLY, THE AUDIO BOLUS BUTTON COVER WAS TORN BUT STILL OPERABLE.

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (CRACKED/DAMAGED CASING) ISSUE. IT WAS ALLEGED THAT THE BATTERY COMPARTMENT WAS DAMAGED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED MALFUNCTION HAS THE ABILITY TO RESULT IN A DELAY IN TREATMENT OR LONG TERM CESSATION IN DELIVERY IF THE DAMAGE IMPACTS THE POWER CIRCUIT OR CARTRIDGE COMPARTMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
743947 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1