ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2016-31160
- Event Type
- Malfunction
- Date Received
- November 10, 2016
- Report Date
- October 18, 2016
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- UDI-DI
- 10840406100167
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1; DATE OF SUBMISSION: 12/09/2016. NO PRODUCT WAS RETURNED. A RETAINED SAMPLE OF THE SAME CARTRIDGE LOT NUMBER U200129 WAS EVALUATED AND TESTED BY PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: A LOT REVIEW INVESTIGATION OF THE INCOMING CARTRIDGES PER LOT WAS COMPLETED PRIOR TO RELEASE OF THIS LOT AND ENSURED THAT ALL CARTRIDGES PASSED FOR THIS LOT. THE RETAINED CARTRIDGE SAMPLE WAS FOUND TO PASS THE VISUAL INSPECTION AND THE FILL, FORCE, AND LEAK TESTING. NO DEFECTS WERE FOUND. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE USER HAD BLOOD GLUCOSE LEVELS OVER 250 MG/DL, BUT LESS THAN 500 MG/DL WITHOUT SYMPTOMS ASSOCIATED WITH A CARTRIDGE ISSUE. THE REPORTER STATED THAT THE USER CHANGED THEIR SITE AND SET. THE USER WAS INSTRUCTED TO CHANGE TO A DIFFERENT LOT NUMBER AND SEE IF THE PROBLEM RESOLVED. THE PATIENTS BLOOD GLUCOSE READINGS DO NOT MEET ANIMAS CRITERIA OF A SERIOUS INJURY. THIS IS POTENTIALLY REPORTABLE BECAUSE THE ISSUE IS NOT ALWAYS OBVIOUS AND DETECTABLE PRIOR TO USE AND CAN RESULT IN UNDER-DELIVERY OF INSULIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 744453 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | 10840406100167 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |