FDA Adverse Event Malfunction Summary report: N

SIMMETRY SYSTEM

MDR report key: 6092998 · Received November 10, 2016

Report

Report Number
3008875054-2016-00022
Event Type
Malfunction
Date Received
November 10, 2016
Date of Event
October 19, 2016
Report Date
November 10, 2016
Manufacturer
ZYGA TECHNOLOGY, INC.
Product Code
LXH
PMA / PMN Number
K151818
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

DURING SIJ FUSION SURGERY A HARDENED AREA WITHIN THE JOINT WAS NOTED DURING THE SECOND DECORTICATION. THEY FORCED THE DECORTICATOR CUTTING ELEMENT THROUGH THE HARDENED AREA CAUSING THE TIP OF THE ELEMENT TO BREAK OFF WITHIN THE JOINT. THE SURGEON DID NOT ATTEMPT TO RETRIEVE THE BROKEN PIECE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
744465 SIMMETRY SYSTEM SIMMETRY LXH ZYGA TECHNOLOGY, INC. 40144 161174

Patients

Seq Age Sex Outcome Treatment
1