IMMULITE 2000 CALCITONIN
Report
- Report Number
- 2432235-2016-00686
- Event Type
- Malfunction
- Date Received
- November 10, 2016
- Date of Event
- October 18, 2016
- Report Date
- December 27, 2016
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED
- Product Code
- JKR
- PMA / PMN Number
- K023304
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE INITIAL MDR 2432235-2016-00686 WAS FILED ON NOVEMBER 10, 2016. ADDITIONAL INFORMATION (12/08/2016): A SIEMENS HEADQUARTERS SUPPORT CENTER SPECIALIST REVIEWED THE DATA PROVIDED BY THE CUSTOMER AND STATED THAT THE CUSTOMER SAW HIGHER RECOVERY WHEN MANUALLY DILUTING A SAMPLE. HOWEVER, THE CUSTOMER DID NOT ADD THE REQUIRED DEIONIZED (DI) WATER TO THE DILUENT BEFORE PERFORMING THE MANUAL DILUTION. THE CUSTOMER HAS AMENDED THEIR STANDARD OPERATING PROCEDURE, SO THAT IF THERE IS EVER A NEED TO PERFORM A MANUAL DILUTION IN THE FUTURE THEY WILL INCLUDE DI WATER AT THE CORRECT RATIO. IMMULITE 2000 CALCITONIN KIT LOT 258 IS PERFORMING AS INTENDED. NO FURTHER EVALUATION OF DEVICE IS REQUIRED.
THE CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER (CCC) SPECIALIST. THE CUSTOMER PERFORMED MANUAL DILUTION FOR CALCITONIN. THE CUSTOMER DID NOT ADD DE-IONIZER TO THE DILUENT PRIOR TO THE DILUTION. THE CAUSE OF THE DISCORDANT CALCITONIN RESULTS ON ONE PATIENT SAMPLE IS UNKNOWN. SIEMENS IS INVESTIGATING THE ISSUE.
DISCORDANT CALCITONIN RESULTS WERE OBTAINED ON ONE PATIENT SAMPLE WHEN RUN NEAT VS. DILUTED ON AN IMMULITE 2000 XPI INSTRUMENT, WHILE USING REAGENT LOT 258. IT IS UNKNOWN WHICH RESULT WAS REPORTED TO THE PHYSICIAN(S). THERE WERE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT CALCITONIN RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 744644 | IMMULITE 2000 CALCITONIN | IMMULITE 2000 CALCITONIN | JKR | SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS, LIMITED | IMMULITE 2000 CALCITONIN | 258 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |