FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 6092243 · Received November 10, 2016

Report

Report Number
2531779-2016-31127
Event Type
Malfunction
Date Received
November 10, 2016
Report Date
October 24, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION: 12/28/2016. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/12/2016 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH A BLANK DISPLAY. THE PUMP WAS OPENED AND THE DISPLAY SCREEN WAS FOUND TO BE CRACKED; A TEST DISPLAY WAS USED AND OPERATED PROPERLY. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (BLANK SCREEN) ISSUE. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY IMPACT THE USER'S ABILITY TO READ SOME OR ALL OF THE INFORMATION ON THE SCREEN WHICH MAY RESULT IN OVER OR UNDER DELIVERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
743405 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1