FDA Adverse Event Malfunction Summary report: N

FREESTYLE PRECISION NEO

MDR report key: 6090219 · Received November 9, 2016

Report

Report Number
2954323-2016-06567
Event Type
Malfunction
Date Received
November 9, 2016
Date of Event
November 2, 2016
Report Date
April 20, 2017
Manufacturer
ABBOTT DIABETES CARE INC
Product Code
NBW
PMA / PMN Number
K140371
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HR
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED METER WAS INVESTIGATED WITH RETAINED TEST STRIPS. METER POWERED ON WITH BUTTON DEPRESSION, BUT NOT TEST STRIP INSERTION. A PORT ISSUE WAS OBSERVED. THE REPORTED COMPLAINT WAS NOT CONFIRMED.

Additional Manufacturer Narrative · 1

THE DEVICE MANUFACTURER DATE FOR THE REPORTED METER IS UNKNOWN. THE DATE ENTERED IS THE DATE ABBOTT DIABETES CARE BECAME AWARE OF THE EVENT. UNLESS FURTHER INFORMATION IS OBTAINED, THIS ISSUE IS CONSIDERED CLOSED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Description of Event or Problem · 1

THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
740183 FREESTYLE PRECISION NEO BLOOD GLUCOSE MONITORING SYSTEM NBW ABBOTT DIABETES CARE INC 71371-75

Patients

Seq Age Sex Outcome Treatment
1