V40 COCR LFIT HEAD 40MM/+8
Report
- Report Number
- 0002249697-2016-03541
- Event Type
- Injury
- Date Received
- November 9, 2016
- Date of Event
- January 24, 2016
- Report Date
- January 26, 2016
- Manufacturer
- STRYKER ORTHOPAEDICS-MAHWAH
- Product Code
- LPH
- PMA / PMN Number
- K061434
- Removal / Correction Number
- 2249697-08/29/2016-007R
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
Narratives
AN EVENT REGARDING ACUTE HIP DISLOCATION INVOLVING A V40 COCR HEAD WAS REPORTED. IT WAS DETERMINED THAT THE HEAD DISASSOCIATED FROM THE STEM. METHOD & RESULTS: DEVICE EVALUATION AND RESULTS: A MATERIAL ANALYSIS HAS BEEN PERFORMED. THE REPORT CONCLUDED: DAMAGES CONSISTENT WITH EXPLANATION DAMAGES AND A LOSS OF TAPER LOCK WERE OBSERVED ON THE ACCOLADE STEM AND V40 HEAD. NO MATERIAL OR MANUFACTURING DEFECTS WERE OBSERVED ON THE SURFACES EXAMINED. MEDICAL RECORDS RECEIVED AND EVALUATION: NOT PERFORMED AS MEDICAL RECORDS WERE NOT PROVIDED. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATES ALL DEVICES ACCEPTED INTO FINAL STOCK MET SPECIFICATIONS. COMPLAINT HISTORY REVIEW: THERE HAVE BEEN NO OTHER EVENTS FOR THIS LOT. CONCLUSIONS: THE SUBJECT DEVICE HAS BEEN IDENTIFIED TO BE WITHIN SCOPE A CAPA . LOT SPECIFIC VOLUNTARY RECALL RA 2016-028 WAS INITIATED FOR THE LFIT V40 COCR HEADS UNDER THE SCOPE OF THIS CAPA. THE INVESTIGATION REVEALED THAT ONLY SPECIFIC CATALOG NUMBERS AND SPECIFIC LOTS ARE IMPACTED BY THE REGULATORY ACTION AND THAT THE AFFECTED LOTS WERE MANUFACTURED ON OR BEFORE MARCH 4, 2011. THE ROOT CAUSE ANALYSIS IDENTIFIED A PROCESS RELATED ANOMALY AS TO THE AFFECTED SIZES AND LOTS. THE AFFECTED PRODUCT HAS ALL BEEN IMPLANTED AND/OR EXPIRED. NO FURTHER INVESTIGATION IS REQUIRED AT THIS TIME.
PATIENT PRESENTED WITH ACUTE HIP DISLOCATION. FEMORAL HEAD, STEM AND INSERT WERE REMOVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 740242 | V40 COCR LFIT HEAD 40MM/+8 | PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED | LPH | STRYKER ORTHOPAEDICS-MAHWAH | LN8MMA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other| R |