FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 6089151 · Received November 9, 2016

Report

Report Number
2531779-2016-30965
Event Type
Malfunction
Date Received
November 9, 2016
Report Date
October 18, 2016
Manufacturer
ANIMAS CORPORATION
Product Code
OYC
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 01/03/2017 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/10/2016 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE BLACK BOX HISTORY SHOWED MULTIPLE CALL SERVICE 087 ALARMS. THE ISSUE WAS DUPLICATED DURING REWIND. THE CALL SERVICE 069 FAILURE WAS WRITTEN TO THE BLACK BOX AS A CALL SERVICE 087 FAILURE. INVESTIGATION REVEALED THAT A LANGUAGE CORRUPTION OCCURRED AT A COMPONENT ON THE PRINTED CIRCUIT BOARD RESULTING IN A CALL SERVICE ALARM. UNRELATED TO THE ORIGINAL COMPLAINT, THE BATTERY COMPARTMENT WAS CRACKED AT THE CASE SEAL BELOW THE BUMPER. ADDITIONALLY, THE AUDIO BOLUS BUTTON COVER WAS DAMAGED/PUNCTURED BUT WAS RESPONSIVE DURING INVESTIGATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(4).

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2016, THE REPORTER CONTACTED ANIMAS ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. IT WAS REPORTED THAT THE PUMP EMITTED A CS 069 CALL SERVICE ALARM. THERE IS NO INDICATION THAT THE PRODUCT ISSUE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ISSUE MAY RESULT IN A LONG TERM CESSATION OF INSULIN DELIVERY IF THE USER IS UNABLE TO RESOLVE THE ALARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
741103 ANIMAS VIBE INSULIN INFUSION PUMP OYC ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1