ANSPACH LONG-HD ATTACHMENT
Report
- Report Number
- 3005985723-2016-00369
- Event Type
- Malfunction
- Date Received
- November 8, 2016
- Date of Event
- November 1, 2016
- Report Date
- November 2, 2016
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
REPORTED EVENT: THE REPORTED DEVICE WAS CONFIRMED TO BE A ANSPACH HD LONG ATTACHMENT, CATALOG # 110920, LOT# J25310715420, RMA #: 235087. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS INDICATE (B)(4) DEVICES WERE INSPECTED AND ACCEPTED INTO FINAL STOCK ON 7/11/2015. DEVICE EVALUATION AND RESULTS: NO FAILURE OF THE ATTACHMENT WAS REPORTED. HOWEVER, A FUNCTIONAL EVALUATION WAS PERFORMED. THERE IS A COMPONENT THAT IS BLOCKING PATH OF THE BURR, PREVENTING INSERTION OF THE BURR. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS RELATED TO P/N 110920, LOT NUMBER J25310715420 SHOWS NO ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. CONCLUSIONS: NO FAILURE OF THIS DEVICE WAS REPORTED (IT WAS RETURNED WITH ANOTHER REPORTED DEVICE). HOWEVER, DURING AN EVALUATION OF THE DEVICE, A BLOCKAGE OF THE BURR PATH WAS FOUND. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
THE SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) WHEN THE BLACK MAGNET FELL OFF, CAUSING THE HANDPIECE TO STOP WORKING.
THE SURGEON WAS PERFORMING A PARTIAL KNEE ARTHROPLASTY USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) WHEN THE BLACK MAGNET FELL OFF, CAUSING THE HANDPIECE TO STOP WORKING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 737324 | ANSPACH LONG-HD ATTACHMENT | STEREOTACTIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | J25310715420 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |